Service manual
Reducing Vaporization Mercury Analyzer Mercury RA-4500
22
(27) Attach the “COLOR” connector.
(28) Attach the cable to the cable clamps. Attach the movable cable such that it moves with
its upper part used as a fulcrum. Fix the cable in a direction where the cable is not
caught by the motor or shaft at the time of vertical driving.
(29) Take care that any cable or the like does
not protrude from the chassis.
(30) When the pump connector has been
disconnected, reconnect it.