1.3. Handling precautions 

Warning

 

  Do not directly measure the sample decomposed with nitric acid. The 

activated carbon in the exhaust gas filter may generate heat. Dilute the 

sample until the concentration of nitric acid falls to 5wt% or less before 

the measurement. 

Caution

 

 

Do not disassemble or modify the product. This may cause a 

malfunction. 

 

Use the product after consulting the manual.    Improper operation 

may cause malfunction or bodily injury. 

 

Comply with the range of specifications. Failure may occur with 

misuse. 

 

Do not operate with wet hands. 

 

Be careful when touching or handling any of the glass components or 
they could be broken. 

 

Be sure to discharge any liquid from the drain tank once the work has 

been completed. It is very important to confirm that the power has 

been turned OFF before starting the discharge. 

 

If y

ou don’t turn-on MA-3000 for a long time, the sample heating tube 

might be obstructed or break by moisture absorption. If you don’t 

turn-on MA-3000 more than 2 weeks, please turn on the MA-3000 

power switch and do conditioning, or dismount the sample heating 

tube and store it in a desiccator to prevent moisture absorption. 

 

After washing the sample boat, first dry it at about 100

C. Then, heat 

at a high temperature to lower the blank. Heating a wet boat to high 

temperatures can damage the sample boat. 

 

1.4. Maintenance precautions 

Warning

 

  Maintenance of the heater must be carried out after the power has 

been turned OFF and the heater has cooled down to close to normal 

room temperature or it could result in burns. 

Caution

 

  Shut OFF the power supply before staring the maintenance work. 

Otherwise electric shock may occur. 

  Be sure to do degreasing cleaning the O-ring before use. If degreasing 

cleaning is not carried out, there is a possibility that the sample heating 

tube will come off during measurement. 

Summary of Contents for MERCURY/MA-3000

Page 1: ...MERCURY MA 3000 Simplified Operation manual NIC 600 2233 06 Nippon Instruments Corporation...

Page 2: ...hat incorrect handling could result in death or severe injury to the operator Caution Indicates that incorrect handling could result in moderate or minor injury to the operator or damage to property 1...

Page 3: ...the discharge If you don t turn on MA 3000 for a long time the sample heating tube might be obstructed or break by moisture absorption If you don t turn on MA 3000 more than 2 weeks please turn on th...

Page 4: ...essage will be displayed if abnormal condition is observed Software will be terminated when OK is clicked Turn off MA 3000 and then fixing them 4 After the check is completed heat conditioning start H...

Page 5: ...method for STD The default method is already set for standard solution but selects the proper method for using SRM Standard Reference Materials Select Flow rate 0 1L min High sensitivity 1pg 10ng 0 2L...

Page 6: ...Solid or Liquid Select a sample feature This is not related to Measuring method Calibration curve Usually please select y ax blank 7 Measurements 7 1 Registration of standard Click the tab of STD 1 C...

Page 7: ...Click the icon and switch to the Analyze screen Review the calibration curve in CALIB tab When some abnormalities are observed on a calibration curve all values of the calibration become red and WARNI...

Page 8: ...e B or L cysteine solution not to vaporize during standby in the sample changer Add small amount of additive B in the boat bottom or add 10 L of 100ppm L cysteine solution When you use additive B plea...

Page 9: ...Standard solution Liquid sample with additive B STD 200 L 1 Biological sample Solid Fish tissue plant tissue ORGANISM SOLID 200 mg 2 Soil sediment sludge SOIL SLUDGE 200 mg 3 Urine Blood URINE BLOOD...

Page 10: ...t of sample or to change the heating conditions Inject 200 L of Distilled Water for several times if high concentration samples are measured Inject 200 L of Standard Solution repeatedly if the sensiti...

Page 11: ...turn on MA 3000 for a long time the sample heating tube might be obstructed by moisture absorption If you don t turn on MA 3000 more than 2 weeks please turn on the MA 3000 power switch and do conditi...

Page 12: ...33 06 11 Nippon Instruments Corporation 14 8 Akaoji cho Takatsuki shi Osaka 569 1146 Japan Please contact the distributor or the following Technical Service E mail info nic rigaku co jp URL https www...

Reviews: