NI PXIe-4300 and TB-4300/B/C
|
© National Instruments
|
7
Figure 4.
TB-4300/B/C Circuit Board Parts Locator Diagram
6.
Tighten the terminal screws to a torque of 0.25 N
·
m (2.21 lb · in.).
Caution
Any wires connected to the ground must be sufficiently insulated from
high voltage.
7.
Use the ground lugs to attach a shield wire to the ground.
Note
Refer to the
NI PXIe-4300 User Manual
for details about shielding the
signals.
8.
Reinstall the strain-relief bar and tighten the strain-relief screws.
9.
Use tie wraps to connect the signal wires to the tie-wrap holes for additional strain relief
when necessary.
10. Reinstall the top cover and tighten the captive top cover screws.
Note
For information about sensors, go to
ni.com/sensors
.