NI PXIe-4300 and TB-4300/B/C
|
© National Instruments
|
5
8.
Secure the module to the chassis using the front-panel captive screws, shown in Figure 2.
Tighten the screws to 0.31 N
·
m (2.7 lb · in.).
Figure 2.
Installing NI PXIe-4300 Modules
Connect the Signals
Note
Refer to the
Read Me First: Safety and Electromagnetic Compatibility
before
removing equipment covers or connecting or disconnecting any signal wires.
To connect signals to the terminal block, refer to Figures 3 and 4 while completing the following
steps:
Note
You can find the pinout names and locations in MAX at any time by
right-clicking the device name under
Devices and Interfaces
and selecting
Device Pinouts
.
1.
Loosen the captive top cover screws and remove the top cover.
2.
Loosen the strain-relief screws and remove the strain-relief bar.
3.
Prepare the signal wire by stripping the insulation no more than 7 mm (0.28 in.).
4.
Run the signal wires through the strain-relief opening.
1
Captive Screws
NI PXIe-1062Q
COOLING CLEARANCE AND F
AN FIL
TER MAINTENANCE REQ
UIRED
.
S
EE MANU
AL.
CO
OLIN
G C
LEA
RAN
CE A
ND
FA
N F
ILTE
R M
AIN
TEN
ANC
E R
EQ
UIR
ED.
S
EE M
ANU
AL.
1