2-2
|
ni.com
Chapter 2
Installation and Configuration
Chassis Cooling Considerations
The NI PXIe-1082 chassis is designed to operate on a bench or in an instrument rack. Regardless
of the configuration you must provide the cooling clearances as outlined in the following
sections.
Providing Adequate Clearance
The primary cooling exhaust vent for the NI PXIe-1082 is on the top of the chassis. The primary
intake vent is on the rear of the chassis where the air is filtered as it enters the power supply
shuttle. The secondary intake and exhaust vents are located along the sides of the chassis.
Adequate clearance between the chassis and surrounding equipment or blockages must be
maintained to ensure proper cooling of the chassis power supply as well as the modules plugged
into the chassis. These clearances are outlined in Figure 2-1. The vent locations for the
NI PXIe-1082 chassis are shown in Figure 2-2. Failure to provide these clearances may result in
thermal-related failures in the chassis or modules.
Figure 2-1.
NI PXIe-1082 Cooling Clearances
1.75 in.
(44.45 mm)
1.75 in.
(44.45 mm)
3
.00 in.
(76.20 mm)
NI PXIe-1082
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com