8.1 Error Messages and Error Handling
DP Series
225
8.1.5 Error Message List
Error messages are listed in Table 8-2.
Table 8-2 Error Message List
ID Message Protection
type
Cause or Necessary action
2
Invalid in This Output
Mode
Message
display only
This output mode does not allow the setting
(execution). Change the output mode.
3
Invalid with Output ON
Message
display only
The output on state does not allow the
setting (execution). Turn the output off, then
set (execute).
4
Invalid with Output OFF
Message
display only
The output off state does not allow the
setting (execution). Turn the output on, then
set (execute).
5 Busy
Message
display only
The busy state does not allow the setting
(execution). Execute after the busy icon
disappears.
6
Invalid with Remote
Control
Message
display only
The remote control state does not allow the
setting (execution). Bring the state to local
control, then set (execute).
10 Sync
Frequency
Error
Message
display only
The external synchronization signal
frequency is out of the range, which does
not allow the setting (execution). Enter the
signal frequency within the range.
13
Auto Cal Disabled
Message
display only
The Autocal setting becomes off because of
out of the calibration range.
15 AGC
Disabled
Message
display only
The AGC setting becomes off because of
out of the calibration range.
20 Invalid
Message
display only
The condition cannot be set (executed).
24 USB
Memory
Unconnected
Message
display only
No USB memory is connected. Connect a
USB memory.
25
USB Memory Removed
illegally
Message
display only
The USB memory was removed before the
eject operation. Perform the eject operation
before removing a USB memory.
26
USB Memory Access
Error
Message
display only
The USB memory has an access error.
27 Too
Many
Files
Message
display only
There are too many files in the USB
memory. Each folder can contain 500 files
or less.
29 Sync
Frequency
Unlocked Message
display only
The output cannot be turned on because of
not locking to the synchronization
frequency.
30
Remote Controller Error
Message
display only
Communication with the remote controller
cannot be made. Make sure the remote
controller cable is connected properly.
31
Internal Memory Error
Message
display only
The internal memory has an error.
Summary of Contents for DP015S
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Page 4: ......
Page 18: ......
Page 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Page 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Page 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Page 298: ......
Page 300: ......
Page 302: ......
Page 303: ......