General cautions

-

  Read this manual before use.

-

  This machine should be used by professionals only.

-

  The built-in safety mechanism should not be re-

moved or modified.

-

  Do not allow unauthorized persons to repair the 

machine.

-

 

Do not use this machine for any other kind of mate-

rial than NextDent

TM 

3D printing Materials.

Package content

-

  NextDent

TM

 LC-3DPrint Box

AC Power cable (250V,10/16A)

6  pcs of UV18W lamps-71 colour

6 pcs of UV18W lamps-78 colour

Metal plate with glass insert

Instruction for use

Electric diagram

Declaration of conformity

Warranty conditions

-

 

-

 

-

 

-

 

-

 

-

 

-

 

-

Transportation / preparing the 

site of machine

-

  For units weighing more than 18kg (units weight: 

22kg), transportation unit like a pallet truck is 

needed and at  least two persons are needed for 

transportation and installation. 

The surface should be flat and free of dust.

Make sure there is enough space on all sides.

Make sure that there is enough space between fan 

and wall.  

This unit is for indoor use only.

 

 

-

 

-

 

-

-

Environmental conditions

-

  Temperature and Humidity: 5~40°C / 80%

Indoor use only

-

 

Steps for installation

-

  Please read instructions for use .

Specifically read general caution including safety 

information. 

Connect all the lamps (12pcs) to all lamp sockets 

which are placed in the cover and the chamber.

The  recommended position of the supplied lamps 

(6pcs of 18W lamps-71color and  6pcs  of 18W 

lamps-78color) is as shown in picture on the right.

Please make sure of rated voltage and power sup-

ply (110-240V 50/60Hz). Only the supplied power 

cable can be used. 

Connect the power cable with connector plug to 

the power socket.

-

 

-

 

-

 

 

 

-

 

-

 

Safety warning label and symbol 

information & Safety information

-

-

-

Do not touch live electric parts.

Turn off power source and disconnect 

power cable before installing or servicing 

the machine. 

Turn off the machine when not in use.

Do not use worn and damaged cables.

Use only well maintained equipment.

 

- Keep away from children. 

- Always turn off the machine before cleaning.

- Do not open the cover when machine is in use.      

- Do not cover the fan at the back.

- Only use the machine with Figure 4

TM

, Fab-

Pro

TM

, and NextDent

TM

 3D printing materials.

-

-

-

-

-

-

-

-

-

-

Wear UV protective glasses.

Do not touch the cover during operation, 

this cover warms up.

Wear protective gloves.

The longest program is for 30 min. the 

temperature inside the working chamber 

can go up to 80°C.

 

Protective earth.

Wear UV protective glasses.

Caution! UV light hazard. Avoid eye and 

skin exposure. Wear proper eye and skin 

protection.

Wear protective gloves.

Be aware of the potential dangers of lifting 

heavy loads. Ignoring this can lead to 

injuries.

This device may only be lifted and moved 

to its final destination for installation by 

two people.

Summary of Contents for LC-3DPRINT BOX

Page 1: ...following instruction is for professionals who want to use the NextDentTM LC 3DPrint Box as a light polymerisation unit post curing oven LC 3DPrint Box is used to dry and polymerize 3D printing object...

Page 2: ...Page 2 of 10...

Page 3: ...pecific voltage Please see the Voltage Chart below to identify your country specific voltage Note If your country is not listed choose the correct voltage from the pulldown list on the print box see v...

Page 4: ...age has changed Continue to press the P button until your country specific voltage rating appears 3 To save the voltage turn off the LC 3D Print box using the power switch on right side of the print b...

Page 5: ...aced in the cover and the chamber The recommended position of the supplied lamps 6pcs of 18W lamps 71color and 6pcs of 18W lamps 78color is as shown in picture on the right Please make sure of rated v...

Page 6: ...he machine is equipped with a fan to cool down the inside and maintain a proper working temperature inside the machine Fan operates when the machine is in use and cool ing is needed If the fan does no...

Page 7: ...essing the start stop button The time will count down and the dots will be blinking By pressing the start stop button the operation time will be memorized automat ically Stop the machine by pressing t...

Page 8: ...structions for use Error message by automatic self checking system The message OPEN will ap pear on the display when a user opens the cover during operation At the same time the machine beeps three ti...

Page 9: ...Page 9 of 10...

Page 10: ...39 49 6151 357 499 Weekdays during normal business hours or by email at Support EMEA 3DSystems com Warranty Hotline from Asia Pacific 852 8191 2453 Weekdays during normal business hours or by email at...

Reviews: