ANOMALIES AND ALARM WARNINGS
Error Codes
When the NI400 is in error status, it shows the letter “E” and error codes one after the
other. Following, there are NI400 error codes.
ERROR CODE
DESCRIPTION
E0
General Communication or FTP/MAIL connection Error.
It refers to log events on web server
E1
USB Drive not found.
E2
Error in file copy.
E3
NI400 is already in run state.
E4
It is impossible to start the NI400.
E5
NI400 is already in configuration state.
E6
Iti s impossible to stop the NI400.
The system shows general error codes because it has a simple display and so it would be
difficult to encode each case. The system will log the event/error and it will be visible and
downloadable from the web.
9
9
Icons and power supply alarm thresholds
NI400 with external power supply
Icon showed: plug
Display: L
High alarm threshold: 14V
Low alarm threshold: 10.5V
NI400 with internal battery pack
Icons showed: cell (battery)
Full battery: voltage≥ 9.6V
Medium battery: 8.76V < voltage < 9.6V
Low battery: Voltage ≤ 8.76V
Display:
3 lines: Voltage ≥ 9.6V
2 lines: 8.76V < voltage < 9.6V
1 line: voltage ≤ 8.76V
High alarm threshold: not present
Low alarm threshold: 8.76V