Next Advance PC77 Operator'S Manual Download Page 5

4

BE SAFE.  ALWAYS MAKE SURE THAT THERE IS NO PRESSURE IN THE CELL BEFORE LOOSENING THE 

BOLTS, AND THAT THE BOLTS ARE SECURELY FASTENED BEFORE APPLYING PRESSURE TO THE CELL.  

USERS MUST ALWAYS WEAR PROTECTIVE APPAREL, INCLUDING EYEWEAR AND MUST BE TRAINED IN 

WORKING WITH HIGH PRESSURE.

Operator’s Manual for the Pressure Injection Cell

desired pressure.

     When the injection is finished, close the 

gas tank valve. Then turn the knob on the 

3-way valve so that it points toward the vent. 

This depressurizes the injection cell. Now 

you can loosen the nut over the ferrule and 

remove the capillary, loosen the three bolts, 

remove the top, and remove your sample 

tube.

Use a wrench to tighten the nut, forcing the ferrule 

to squeeze tightly around the capillary.

USE OF ADAPTER FOR 1 ML TUBES

     We supply one adapter with the pressure 

injection cell so that 1 ml tubes are as easy to 

work with as taller tubes. 

Photograph of a 1 ml tube with a hinged cap (left) 

and of a 1 ml tube with the hinge cut off and the 

bottom placed in an adapter composed of soft 

tubing (right).

     The central hole in the base of the pressure 

injection cell is sized so that 1.5 or 2 ml

sample tubes will protrude above the surface, 

making placement and retrieval easy. A 

(surprisingly) simple adapter will enable 1 ml 

microcentrifuge tubes to protrude, too. The 

adapter is a piece of soft tubing (1/4 inch 

inside diameter, 3/8 inch outside diameter). 

As with any tube with a hinged cap, first cut 

off the hinge so that the top of the tube will fit 

in the recess in the top of the pressure 

injection cell. Next, slide the adapter onto the 

bottom of the 1 ml sample tube, as shown in 

the figure below.

STIRPLATE OPERATION

(PC77-MAG model only)

Operation of the stirplate is straightforward. 

The power supply should be plugged in to the 

jack at the rear. The speed control knob is on 

the left. Clockwise rotation increases the 

speed. The power switch is on the right.

MAINTENANCE

Periodically, the Teflon ferrule should be 

replaced. Simply remove the nut covering the 

ferrule, remove the ferrule, place a new 

ferrule, tapered side down into the fitting, and 

replace the nut. Your pressure injection cell 

comes with 10 ferrules (plus one installed in 

the top). Additional ferrules can be purchased 

directly from Chromatography Research 

Supplies (part number 214204, 1/8” to 0.4mm 

PTFE ferrules), at www.chromres.com

DISCONNECTING AND                        

RECONNECTING THE UNIT

Before you disconnect the pressure injection 

cell, mark lines on the nut and the valve, 

close the valves by the pressure source, and 

release any pressure inside the cell by 

opening the 3-way valve. When you recon-

nect the unit, you can use the same ferrule 

and tubing. Simply tighten the nut slightly 

more than last time. You should feel the 

wrench get very hard to turn once you go 

beyond the nut’s previous position.

Summary of Contents for PC77

Page 1: ...Pressure Injection Cell Operator s Manual ...

Page 2: ...close the valves on the pressure source and release any pressure from inside the pressure injection cell before opening the cell removing the top or connecting or disconnecting tubing to the cell Never open the valves or try to pressurize the pressure injection cell without all connections properly tightened and the top securely bolted to the base of the cell Wear protective eyewear Users must be ...

Page 3: ...se the test tubes capillary and the Pressure Cell in accordance with local regulations PARTS OF THE PC77 MAG PRESSURE INJECTION CELL Nut Bolt Top Base Teflon Ferrule under the nut 3 Way Valve Inlet Knob Vent Magnetic Stirplate 2 BE SAFE ALWAYS MAKE SURE THAT THERE IS NO PRESSURE IN THE CELL BEFORE LOOSENING THE BOLTS AND THAT THE BOLTS ARE SECURELY FASTENED BEFORE APPLYING PRESSURE TO THE CELL USE...

Page 4: ...lator to the SETTING UP YOUR PRESSURE INJECTION CELL Setting up the Pressure Injection Cell consists of attaching it to 1 8 inch outside diameter thick walled stainless steel tubing not supplied with the PC77 Pressure Injection Cell First slide the tubing through the inlet nut until it hits a stop Finger tighten the nut and then tighten it with a wrench for 1 1 4 turns Connect the other end of the...

Page 5: ...protrude too The adapter is a piece of soft tubing 1 4 inch inside diameter 3 8 inch outside diameter As with any tube with a hinged cap first cut off the hinge so that the top of the tube will fit in the recess in the top of the pressure injection cell Next slide the adapter onto the bottom of the 1 ml sample tube as shown in the figure below STIRPLATE OPERATION PC77 MAG model only Operation of t...

Page 6: ...ease contact customer service by email at support nextadvance com or by telephone SPECIFICATIONS Max Pressure 17 MPa 2500 psi Size 18 cm 6 5 in long x 10 cm 4 in wide x 19 cm 8 in high Weight 1 6 kg 3 5 lbs Relative Humidity 5 90 non condensing Operating Temperature 4 40 C Altitude 2000m Storage Temperature 40 to 50 C Meets requirements Power Requirement PC77 MAG model 24V DC 0 8 A WARRANTY Next A...

Page 7: ...or implied with respect to its Products NEXT ADVANCE MAKES NO WARRANTY RESPECTING THE MER CHANTABILITY OF THE PRODUCTS OR THEIR SUITABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE OR USE Next Advance shall not be liable for indirect special incidental or consequential damages of any nature Any recovery for any claim shall be limited to the original purchase price for the product WARNINGS AND CAUTION...

Reviews: