Newport NPC DIG Series Instruction Manual Download Page 7

 

 

7

 

Purchased Part Package 

 

Please check the completeness of the delivery after receiving the shipment: 

 

 

piezo amplifier 

NPCxxxDIG

 

 

wide range power supply 24 V DC 

 

RS232 cable 

 

USB cable 

 

instruction manual 

 

CD-ROM with driver, software, and instruction manual 

Instructions for using piezo electrical elements and power supplies 

 

Piezoelectric actuators from 

Newport Corporation

 are controlled by voltages up to 150 V. 

These values can be quite hazardous. Therefore, read the installation instructions carefully 
and ensure that only authorized personnel handle the power supply. 

 

After transportation, piezoelectric actuators should be allowed to adapt to room temperature 
for approximately 2 hours before being switched on. 

 

Piezoelectric actuators are made from ceramic materials with and without metallic casings. 
The  piezo-ceramic  is  a  relatively  brittle  material.  This  should  be  noted  when  handling 
piezoelectric actuators. All piezo elements are sensitive to bending or shock forces. 

 

Due to the piezoelectric effect, piezo actuators can generate electrical charges by changing 
the mechanical load or the temperature, or by actions such as the ones described above. 

 

Piezoelectric actuators are able to work under high compressive forces. Only actuators with 
a pre-load can be used under tensile loads (these tensile forces must be less than the pre-
load  given  in  the  data  sheet).  Please  note  that  acceleration  of  the  ceramic  material  (e.g., 
caused by fall down, discharging, or high dynamic application) will occur. 

 

After  excitation  of  the  actuators  by  a  voltage  in  the  upper  control  range,  the  ceramic  will 
move and generate an opposite high voltage after disconnection. 

 

Heating  of  the  ceramic  material  will  occur  during  dynamic  operation  and  is  caused  by 
structure  conditional  loss  processes.  This  may  cause  failure  if  the  temperature  exceeds 
specified values cited below. With increasing temperature up to the Curie temperature  T

C

(usual values approx.  140°C - 250°C) the piezoelectric effect disappears. We recommend 
working in temperatures up to T

C

/2 (normally up to 80°C). 

 

Piezoelectric  actuators,  such  as  stacks  or  other  devices,  work  electrically  as  a  capacitor. 
These elements are able to store electrical energy over a long period of time (up to some 
days) and the stored energy may be dangerous. 

 

If  the  actuator  remains  connected  to  the  drive  electronics,  it  will  be  unloaded  within  a 
second after shutdown and quickly reaches harmless voltage values. 

 

Piezo  actuators  can  only  generate  voltages  by  warming  or  cooling  (caused  by  the 
longitudinal  change).  The  discharge  potential  should  not  be  ignored  due  to  the  inner 
capacitance. This effect is insignificant at usual room temperature. 

 

Piezo  actuators  from 

Newport  Corporation

  are  adjusted  and  glued.  Any  opening  of  the 

unit  will  cause  misalignment  or  possible  malfunction  and  will  result  in  the  loss  of  the 
guarantee.. 

 

Please only use original parts from 

Newport Corporation

 

Please contact 

Newport Corporation

 or your local representative if there are any problems 

with your actuator or power supply. 

 

Caution! 

Shock forces may damage the built-in ceramic elements. Please avoid such forces, 

and handle the units with care, otherwise the guarantee will be lost. 

Summary of Contents for NPC DIG Series

Page 1: ...n Manual Digital Piezo Amplifier Series NPCxxxDIG Models NPC50DIG NPC300DIG Please read carefully before switching on the power Please see safety instructions for using piezoelectric actuators and pow...

Page 2: ...www newport com 2 English version last change 2017 02 16 by TP...

Page 3: ...3...

Page 4: ...for checking the function of the system quick start 11 8 How to operate the digital amplifier series NPCxxxDIG 12 8 1 Common Introduction 12 8 2 User Elements Connections 12 8 2 1 Front Panel 12 8 2...

Page 5: ...rporation You will also find additional information regarding piezoelectric products Definition All systems from Newport Corporation such as electronics actuators and optical systems are called units...

Page 6: ...uction and form put by us into circulation is in accordance with all relevant essential requirements of the EMC Directive 2014 30 EU This declaration is no longer valid if the device is modified witho...

Page 7: ...l occur After excitation of the actuators by a voltage in the upper control range the ceramic will move and generate an opposite high voltage after disconnection Heating of the ceramic material will o...

Page 8: ...regular waste Instead it is the user responsibility to dispose of waste equipment according to the local laws The separate collection and recycling of the waste equipment at the time of disposal will...

Page 9: ...rs Only work with the units in a clean and dry environment Only specially prepared units e g actuators can work under other conditions Please only use original parts from Newport Corporation Newport C...

Page 10: ...cables e g power cords sensor cables actuator cables optical cables Do not block ventilated openings or place the units near a radiator oven or other heat sources Plug in the power cord completely so...

Page 11: ...r shape should be parallel each to each other and not have any scratches If there is any damage to the system please contact our local representative immediately If the packaging material is damaged p...

Page 12: ...high speed positioning with the highest accuracy available It includes rise time optimization and an active oscillation damping for every special application The NPCxxxDIG comes with an auto calibrati...

Page 13: ...13 8 2 2 Back Panel...

Page 14: ...www newport com 14 8 3 Function The following sketch shows the function groups of the amplifier sketch 1 function groups of the amplifier...

Page 15: ...akes for the output to get a certain value For a step input it is the time taken for the output to equal the input yi yi ki err Ts Ts 1 sample frequency 50 kHz 1 ki time constant yi output of i term I...

Page 16: ...er of magnitude While using open loop you can also switch on the notch filter to reduce stimulation of the resonant frequency Using the notch filter might increase noise The PID specific parameters kp...

Page 17: ...500Hz Actuator measurement connector PIEZO 15pin D Sub plug Monitor output 0 10V programmable signal source Monitor output impedance 1k Modulation input 0 10V disconnectable Modulation input impedanc...

Page 18: ...ription value monsrc value monitor output default 0 0 position value in closed loop 1 reference input 2 PID command voltage 3 closed loop deviation incl sign 4 closed loop deviation absolute 5 actuato...

Page 19: ...n loop position voltage back from the measured monitor voltage use the following formula Umes OL Umon 2 5V 2 The default setting is the position value in closed loop see No 0 sketch 3 monitor output 8...

Page 20: ...is connected to this plug The actuator voltage goes to the piezo the position signal from the measurement system goes to the DSP and the integrated actuator information is stored on an ID chip which g...

Page 21: ...ohm operation time of actuator since shipping minutes rgver version number of loop controller request table 6 commands without values commands with values command value Enter command description value...

Page 22: ...amplitude sine 0 to 100 gosin amplitude offset sine 0 to 100 gfsin generator frequency sine 0 1 to 9999 9 Hz gatri generator amplitude triangle 0 to 100 gotri amplitude offset triangle 0 to 100 gftri...

Page 23: ...rface every 500 milliseconds This value can then be read out via the interface Subsequent increasing of the query sequence for this value does not increase the refreshing rate If the piezoelectric act...

Page 24: ...11 10 9 211 210 29 0 0 0 generator off 0 0 1 sine on 0 1 0 triangle on 0 1 1 rectangle on 1 0 0 noise on 1 0 1 sweep on 0 512 1024 1536 2048 2560 12 212 0 notch filter off 1 notch filter on 0 4096 13...

Page 25: ...of the impulses trglen 1 or decrease the movement speed of the actuator Furthermore you must be sure that the trigger range between trgss and trgse passes through plus 0 2 of the total stroke otherwis...

Page 26: ...xt Trigger position is set to 10 m the trigger impulse will be generated when the 10 m position is reached from a lower position than 10 m rising edge Extended trigger functions trgedge 4 5 7 The func...

Page 27: ...ed This principle depends on whether or not trgss and trgse are bordered by a working area that crosses at least 0 2 of total stroke out of the borders trgss lower trigger position this is the lowest...

Page 28: ...triangle This will require that the relevant functional parameters be set frequency amplitude offset or symmetry The function generator does not need to be activated gfkt 0 To select a scan set the sc...

Page 29: ...combined use of trigger generation and scan function permits a highly exact scan of a probe Acceleration forces and with it oscillations are minimized by using a sine function and actions can be initi...

Page 30: ...OD input the actuator is in the most negative position pos 1 of its deviation pos 1 to pos 4 see sketch 2 By pressing the knob in the operation mode changes to closed loop and the OL CL LED changes to...

Page 31: ...onant frequency Use function generator in the menu of the actuator Set the amplitude to 5 the offset to 0 and the sweep time to 1 1sec decade Set the amplifier to open loop OL CL LED lights green With...

Page 32: ...justments reduce the ki value Reduce the kd value Check whether the function generator is still working read back parameters always deviate at the same level from the digital target value Please switc...

Page 33: ...ng in nm resolution you must warm up the amplifier 2 hours before use A constant temperature is beneficial Consider that a variation of 5 Kelvin results in elongation by 13 m in steel with a length of...

Page 34: ...www newport com 34 12 Your Notes...

Reviews: