Newport iServer MicroServer Operator'S Manual Download Page 6

PART 1

INTRODUCTION

1.1  Safety and EMC Considerations

Refer to the CE Approvals Section.

EMC Considerations

Whenever EMC is an issue, always use shielded cables.

Never run signal and power wires in the same conduit.

Use  twisted-pair wires for signal connections.

Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems persist.

Failure to follow all instructions and warnings may result in injury!

1.2  Before You Begin
Inspecting Your Shipment: 

Remove the packing slip and verify that you have received

everything listed. Inspect the container and equipment for signs of damage as soon as
you receive the shipment. Note any evidence of rough handling in transit. Immediately
report any damage to the shipping agent. The carrier will not honor damage claims
unless all shipping material is saved for inspection. After examining and removing the
contents, save the packing material and carton in the event reshipment is necessary.

Customer Service:  

If you need assistance, please contact the Customer Service

Department nearest you.

Manuals, Software:

The latest Operation Manual as well as free software and iServer

Mail Notifier are available at 

the website listed on the cover page of this manual or

on the CD-ROM enclosed with your shipment

.

1.3  Description
View Tempe Humidity with a Web Browser.  

The iServer let’s you monitor and

record Temperature and Relative Humidity over an Ethernet network or the Internet with
no special software except a Web browser. The iServer serves Active Web Pages to
display real time readings and display charts of temperature and humidity, or log data in
standard data formats for use in a spreadsheet or data acquisition program such as
Excel or Visual Basic.

Adjustable Charts.

The Java™ Applet chart scales are fully adjustable on the fly. For

example, the chart can display one day, one week, one month or one year. Temperature
and humidity can be charted across the full span (0-70ºC, and 0-100% RH) or within any
narrow range (such as 20-30ºC).

2

Summary of Contents for iServer MicroServer

Page 1: ...http www newportUS com iServer NEWPORT Electronics Inc Temperature Humidity Operator s Manual...

Page 2: ...ate device upon certification The information contained in this document is believed to be correct but NEWPORT Electronics Inc accepts no liability for any errors it contains and reserves the right to...

Page 3: ...ng Sensor 15 4 3 1 1 Java Runtime Environment 1 4 Setup Instructions 15 4 3 1 2 Java Runtime Environment 1 5 5 0 Setup Instructions 16 4 3 1 3 Browser Proxy Selection 16 4 3 2 Adjustable Chart 17 4 3...

Page 4: ...ning an IP Address using iCONNECT 12 Figure 4 3 Accessing the iServer s Home Page 12 Figure 4 4 Access Control Menu 13 Figure 4 5 iServer Home Page Menu 14 Figure 4 6 Login Password 14 Figure 4 7 Read...

Page 5: ...about the risk of electrical shock CAUTION Risk of danger Tells you of circumstances or practices that can effect the instrument s functionality and must refer to accompanying documents TIP Provides...

Page 6: ...nd carton in the event reshipment is necessary Customer Service If you need assistance please contact the Customer Service Department nearest you Manuals Software The latest Operation Manual as well a...

Page 7: ...complete with one temperature humidity wand probe full documentation and software Alarms and Email If temperature or humidity exceeded a set limit the iServer can trigger an alarm that could be sent b...

Page 8: ...ion DIP Switch Usage 1 N C not used 2 To change to default factory settings 3 To enable disable DHCP 4 N C not used To set the iServer to factory default settings slide DIP switch 2 to ON position Pow...

Page 9: ...d C F Button Change display units of measurement from C to F TIME Button Change display from DATE and TIME to RH and TEMP STBY Button Will stop the Flash from recording BKLT Button Display Back Light...

Page 10: ...D Backlight and iServer Board will be on and running on the battery power Refer to Section 2 7 Disconnect the power supply before proceeding Remove Cover as shown Figure 2 4 Opening the Unit MOUNTING...

Page 11: ...need to make sure that the LCD PWR field on the Configuration page of the iServer is set to UPS see Section 4 3 5 F Battery installed and S5 jumper in storage position Factory Default if there is a p...

Page 12: ...ddress is your computer s unique hardware number When you re connected to the LAN from your computer a correspondence table relates your IP address to your computer s physical MAC address The MAC addr...

Page 13: ...ividual computers and devices to be recognized over a network based on a specific name instead of an IP address For example instead of having to use http 128 100 101 254 IP address you would use only...

Page 14: ...cable between your computer and the iServer and modify the IP address or any other settings within the iServer 3 6 Port Number All TCP connections are defined by the IP address and a port number A por...

Page 15: ...ing the iServer directly to your network you can configure your PC s network connection with an IP address 128 100 x x that is in the same range as the iServer s default IP address 128 100 101 254 and...

Page 16: ...owser to access the iServer s web pages Consult with your IT department for obtaining an IP address Figure 4 2 Assigning an IP Address using iCONNECT d To access the iServer for Configuration Click on...

Page 17: ...onses back Figure 4 1 you can go to a Web browser and type in http 128 100 101 254 and it will take you to the iServer s main web page Select Access Control button you ll be asked for the password The...

Page 18: ...mpted for a password as shown in Figure 4 5 Figure 4 6 Login Password There are 2 different access levels 1 Admin Password administrator allows certain groups and individual users to access and modify...

Page 19: ...alled and configured according to the following instructions If you do not have Java Runtime Environment you may download it from our website or contact the Customer Service Department nearest you Fig...

Page 20: ...ad Sensor web page there may be a problem with access through a winsock proxy server In such cases your network administrator will need to provide the winsock proxy server and port s If the administra...

Page 21: ...ontact the Customer Service Department nearest you Figure 4 8 Adjustable Chart Chart Flash Archive Main Menu http 128 100 101 254 Title 1 Hours Div 1 Day 1 Day 1 Week 1 Month 1 Year 1 Minute 1 Hour 02...

Page 22: ...gram to obtain the screen shown in Figure 4 9 IP Address Place the IP address of the iServer from which you want to download data in this window Output Filename Assign the path and the file name in wh...

Page 23: ...net Protocol address is a 32 bit number that identifies each sender or receiver of information that is sent in packets across the Ethernet or the Internet The iServer s default IP address is 128 100 1...

Page 24: ...ime to review verify that the setup took place as it was intended 2 RTC is set and recording is in progress The only possible setups are to update Date and Time of the Real Time Clock along with check...

Page 25: ...11 Configuration Device Setup Menu Below are the definitions of terms used in the Configuration Device Setup Menu A Device Setup Clicking on the No 1 2 or 3 allows you to modify the Sensor Parameters...

Page 26: ...nnected or the voltage is low the statement Low Battery will be added to the above statements You may also see a statement Archive Busy if another client is busy receiving archive data In this case re...

Page 27: ...the Flash is full then the iServer will continue to record and overwrite the existing data Nonwrap You only need to define the Start Date and Time The iServer will record data until the Flash is full...

Page 28: ...the Update button if you then need to make any changes in the Flash Recording section you must either select StopRecord followed by clicking the Update button or push the reset button on the side of t...

Page 29: ...r product if a computer will be obtaining data concurrently with or instead of using this iServer s internal charting process H Remote Access Remote IP Address iServer can establish a connection to a...

Page 30: ...he iServer sends CR after each temperature humidity and dewpoint value This will be done either in Continuous or Command mode This is how the data will appear on the host T_75 70F H_37 90 D_44 90F If...

Page 31: ...r product Generally httpget is used for simply programming an IP address to the iServer or for quickly obtaining a reading of from a device The iServer product must be configured from the configuratio...

Page 32: ...hen installing the Mail Notifier software 1 Open up a command window or have a DOS window a Click on start menu b Click on Run c In the dialog box that appears type cmd or command and click on OK butt...

Page 33: ...06 b6 Then use the following command to assign new IP to the device Httpget r S 00000000 128 100 101 254 1 where 0000000 is admin password If the password is wrong the unit will ignore the new IP If t...

Page 34: ...a plus IP address Use this command to view ARP cache entries associated with one particular interface on a network with multiple adapters arp g Same as arp a arp N Use this command to display ARP ent...

Page 35: ...of alarm conditions alarm conditions can be monitored on a network isolated from the internet and forwarded to connections on the Internet The Mail Notifier utility operates under Windows 98 NT 4 0 20...

Page 36: ...ss The following instructions work with some versions of Microsoft Outlook Note though that some email systems may prevent the use of Mail Notifier due to security constraints The simplest way to obta...

Page 37: ...obtained from the device Figure 4 17 iServer Mail Notifier Device Setting 4 8 RH Temp Excel Software The RH Temp Excel logging software enables you to use Microsoft Excel for logging graphing and sto...

Page 38: ...upgradeable Including an Excel program for automatic data logging within definable time intervals compatible with all Windows operating systems 34 PART 5 SPECIFICATIONS SENSOR SPECIFICATIONS Relative...

Page 39: ...st Name eis and Last 4 digits from the MAC address Login Password 12345678 Admin Password 00000000 DHCP Disabled Flow Control None End Character 0D Hex Carridge Return Terminal Server Server Type Comm...

Page 40: ...32 bit pattern of bits used to determine which part of the IP address is the network portion and which part is the host portion MAC Media Access Control Address is your computer s unique hardware numb...

Page 41: ...pport approximately 216 65 536 computers per network The IP addresses range in binary from 10000000 00000000 xxxxxxxx xxxxxxxx to 10111111 11111111 xxxxxxxx xxxxxxxx The IP addresses range in decimal...

Page 42: ...Mask Dotted Notation Mask Binary Mask Bits Class A 255 0 0 0 Default 0 255 192 0 0 2 255 224 0 0 3 255 240 0 0 4 255 248 0 0 5 255 252 0 0 6 255 254 0 0 7 255 255 0 0 8 255 255 128 0 9 255 255 192 0...

Page 43: ...01010010 DC3 19 13 00010011 S 83 53 01010011 DC4 20 14 00010100 T 84 54 01010100 NAK 21 15 00010101 U 85 55 01010101 SYN 22 16 00010110 V 86 56 01010110 ETB 23 17 00010111 W 87 57 01010111 CAN 24 18...

Page 44: ...11 Crtl Q Data Control 1 XON SOH 01 01 Crtl A Start of DC2 18 12 Crtl R Data Control 2 Header STX 02 02 Crtl B Start of Text DC3 19 13 Crtl S Data Control 3 XOFF ETX 03 03 Crtl C End of Text DC4 20 1...

Page 45: ...ervoltage Note The ac dc power adaptor must have Safety Qualified Agency Approvals for CE with Double Insulation rating The ac dc power adaptor is 9Vdc The minimum output current rating is 500mA EMC E...

Page 46: ...l of the published electrical mechanical and operational performance characteristics All tests and calibrations were performed with instruments equipment and standards which are traceable to the U S N...

Page 47: ...rwise shall not exceed the purchase price of the component upon which liability is based In no event shall NEWPORT be liable for consequential incidental or special damages CONDITIONS Equipment sold b...

Page 48: ...newportuk co uk e mail sales newportuk co uk Newport Electronics B V Postbus 8034 1180 LA Amstelveen The Netherlands TEL 31 20 3472121 FAX 31 20 6434643 Toll Free 0800 0993344 www newport nl e mail i...

Reviews: