20
2.
Connect the thermocouple calibrator output signal to
the thermocouple input connector P3 as shown in
Figure 4-2. Connect the negative (red) thermocouple
wire to connector P3, Pin 1. Connect the positive
thermocouple wire to connector P3, Pin 2. Be sure to
observe polarity and use the correct thermocouple
wire (Type J wire for Type J meter).
Figure 4-2. Calibration a Thermocouple Calibrate
3.
Connect the power to the meter. Allow 30 minute
warm up before calibration of 0.1° resolution meters.
Refer to the appropriate model number in
Table 4-1
for the proper calibration points for your particular
meter.
METER
P3
P2
P1
T/C WIRE
AC
T/C CALIBRATOR
0
°
C
9
Figure 3-3. Installation Information
PANEL CUTOUT
PANEL CUTOUT
CASE
0.25 [6.4] MAX
0.03 [0.8] MIN
PANEL THICKNESS
0.874 +.012/-.000
[22.20 +0.30/-0.00]
R 0.06 [1.5]
4 PLCS
[68.00 +0.70/-0.00]
2.677 +0.028/-.000
MOUNTING
SCREW HOLES
MOUNTING SCREW (2)
MOUNTING CLAMP
LENS
CLAMP
CLAMP
SCREWS
S1
S
Z
20
2.
Connect the thermocouple calibrator output signal to
the thermocouple input connector P3 as shown in
Figure 4-2. Connect the negative (red) thermocouple
wire to connector P3, Pin 1. Connect the positive
thermocouple wire to connector P3, Pin 2. Be sure to
observe polarity and use the correct thermocouple
wire (Type J wire for Type J meter).
Figure 4-2. Calibration a Thermocouple Calibrate
3.
Connect the power to the meter. Allow 30 minute
warm up before calibration of 0.1° resolution meters.
Refer to the appropriate model number in
Table 4-1
for the proper calibration points for your particular
meter.
METER
P3
P2
P1
T/C WIRE
AC
T/C CALIBRATOR
0
°
C
9
Figure 3-3. Installation Information
PANEL CUTOUT
PANEL CUTOUT
CASE
0.25 [6.4] MAX
0.03 [0.8] MIN
PANEL THICKNESS
0.874 +.012/-.000
[22.20 +0.30/-0.00]
R 0.06 [1.5]
4 PLCS
[68.00 +0.70/-0.00]
2.677 +0.028/-.000
MOUNTING
SCREW HOLES
MOUNTING SCREW (2)
MOUNTING CLAMP
LENS
CLAMP
CLAMP
SCREWS
S1
S
Z