Swept-Wavelength Tunable Laser Source
™
Computer Control • 71
Set Starting Wavelength for Wavelength Scans
Syntax
[:SOURce]:WAVElength:STARt <val>
Description
Sets the starting wavelength for wavelength scans.
Note: If the wavelength ramp is enabled, the laser will reset
to 2 nm before the start wavelength. See page 26 for more
on the ramp, or page 76 for information on enabling or
disabling the ramp.
Argument
x.xxx
Range: The value must be less than the stop wave-
length and must be within the laser’s operating
range. (The minimum and maximum values are
laser dependent. See “Query Wavelength” on
page 65.)
Units: nanometers (nm).
Example
Set scans to start at 1530 nm.
:WAVE:STAR 1530
⇒
OK
Query Starting Wavelength
Syntax
[:SOURce]:WAVElength:STARt ?
Description
Queries the starting wavelength for scans.
Note: If the wavelength ramp is enabled, scans will start 2
nm before this wavelength. See page 26 for more on the
ramp, or page 76 for information on enabling or disabling
the ramp.
Response
“x.xxxx”
Units: nanometers (nm).
Example
:WAVE:STAR ?
⇒
1530.0000
(Scans will start at 1530 nm, or at 1528 nm if the
ramp is enabled.)