23
-
③
The Power lamp lights up again and the auto
zero adjustment value is reset.
④
Make sure that there is no gas around the
gas detector. Press the enter key and
simultaneously turn the zero adjustment
volume (ZERO) to set the indicated value on
zero.
⑤
Set the Mode switch at zero to return to
normal mode.
2) Analog span adjustment
①
Set the Mode switch at 2 to select maintenance
mode 2.
②
Press down the reset key until the Power lamp
goes off.
POWER
ALARM
2
1
RESET
BZSTOP
BACKUP
On
POWER
ALARM
2
1
RESET
BZSTOP
BACKUP
Adjust
to zero.
POWER
ALARM
2
1
RESET
BZSTOP
BACKUP
Check to see if
the indicator
reads 0.
Always perform analog zero adjustment before doing analog span adjustment. Otherwise the span cannot be
adjusted properly when the analog span adjustment is performed.
CAUTION
POWER
ALARM
2
1
RESET
BZ STOP
BACK UP
POWER
ALARM
2
1
RESET
BZ STOP
BACK UP
Off