27
8.5 D
ISPLAYING THE
C
URRENT
V
ALUES
The current value reading for the
FOAM
and
HIGH FOAM
sensors is displayed
through the LED on the front panel of the BioFlo 2000 Fermentor. Perform the
following steps to display these signals:
8.5.1
On the front panel of the BioFlo 2000 Fermentor press and release the
MODE SELECT KEY
to scroll through the modes. Scroll through
the select modes until the light next to
FOAM
illuminates red.
New Brunswick Scientific Co., Inc.
Bio Flo 2000
FERMENTOR
ZERO
CONT
SET
LEV
FOAM
SPAN
DO
TEMP
AGIT
NUTR
HARV
pH
MODE
SELECT
FUNCTN
SELECT
D.O.
CONTROL
AGIT
ON/OFF
NEW BRUNSWICK SCIENTIFIC
Edison NJ USA
BioFlo 2000
Front Panel
8.5.2
T
T
he 2 middle digits of the LED shows the current
FOAM
condition.
These values can be interpreted from the following table:
Sensor Status Table
Foam Value
Foam Sensor
High Foam Sensor
0 0
No Foam
No Foam
0 1
Foam
No Foam
1 0
No Foam
No Foam
1 1
Foam
Foam