New Age BIOSONYC User Manual Download Page 8

Rev. 02 of 10/12/04 

8/24 

3 . 3

 

–   T u r n   o n   t h e   d e v i c e  

 

In order to turn BIOSONYC on, press the button 1/0 – ON/OFF. The display will show a 
welcome message, with the name and the software version, and the Selection Menu screen 
will appear. 

 

3 . 4

 

–   C h o o s e   t h e   p r o g r a m  

 

On the Menu, choose the program with UP/DOWN keys. 

 

3 . 5

 

–   S t a r t   o f   t h e r a p y  

 

To start the therapy press 

start

; the power of the stimulation is already preset but can be 

set  by  pressing  (+)(-)  keys.  The  minimum  increase  of  power  is  0.1  W/cm2  and  the 
maximum is 3.0 W/cm2. 

 

NOTE:

  if  no  head  has  been  fitted  into  the  device  at  the 

start

,  you  can  hear  an  acoustic 

warning and the display will show an error message until it is inserted, then it will start up 
again automatically. 

 

3 . 6

 

–   S e t t i n g   t h e   p o w e r   o f   e m i s s i o n  

 

Use  the  (+)  and  (–)  keys  to  set  the  intensity  of  emission.  Generally  speaking,  it  stays 
constant throughout the duration of the program; however, in case of pain or overheating 
of the area being treated, you must decrease the power using the (–) key or else break off 
the therapy by pressing 

pause stop

 

NOTE:

  during  the  program  it  is  possible  to  increase  or  decrease  the  Timer  by  pressing 

UP/DOWN keys (one minute each touch).

 

 

3.6.1

 

– Conta ct co ntro l  

 

To detect the contact between head and skin, fasten the bracelet in the patient's wrist and 
in the back output of the device. If the head does not adhere perfectly to the treating area 
or to the mean of transmission (gel, water in immersion) the device will let out an acoustic 
signal and interrupt the program; adjust the contact to continue. 

 

3 . 7

 

–   P r o g r a m   i n t e r r u p t i o n  

 

The stimulation  will stop automatically as soon as the Timer reaches 0 and  you hear  the 
end program acoustic signal; if you want to interrupt the program before the end, press the 

pause stop

 key: once to make a pause (to continue press 

start

), twice to stop the program. 

 

3 . 8

 

–   F r e e   p r o g r a m   s e t u p  

 

Biosonyc allows to create and save up to 5 free programs listed from prog. no 41 to no. 
45.  
To save a free program go to the program list and select one of the mentioned programs.  
Press STOP. You enter the parameter configuration menu PARIMP. From here with the 
keys  + and – adjust the value of the parameter shown, while with the up/down arrow key 
select another parameter. 

Summary of Contents for BIOSONYC

Page 1: ...Telefax 39 0545 369028 Website www newageitalia it e mail info newageitalia it This document is the property of New Age Italia srl All rights reserved It is forbidden to copy or reproduce it in any w...

Page 2: ...hoose the program 8 3 5 Start of therapy 8 3 6 Setting the power of emission 8 3 6 1 Contact control 8 3 7 Program interruption 8 3 8 Turn off the device 9 Chap 4 APPLICATION TECHNIQUES 9 4 1 Direct c...

Page 3: ...ogram 18 Chap 6 POWER SUPPLY 19 Chap 7 MAINTENANCE 19 7 1 Head and cables 19 7 2 Device 19 7 3 Replacing the power supply cable 19 Chap 8 SYMBOLS 20 Chap 9 WARNINGS 21 Chap 10 TECHNICAL SPECIFICATIONS...

Page 4: ...2 W h y u s e t h e B I O S O N Y C BIOSONYC houses inside itself all the technology needed to generate ultrasounds through the use of pre set programs contained in its memory The generator carries o...

Page 5: ...activity however in many cases they may also be due to aging rheumatisms muscular hypotrophy lack of balance due to little movement these cases are increasing steeply due to the progressive rising of...

Page 6: ...in the immediate neighbourhood of the eyes or the uterus abdomen lumbar area in case of pregnancy due to the risk of provoking cavitation even with a proper dosage The presence of articular prostheses...

Page 7: ...PPLY OUTPUT ON OFF DISPLAY START UP DOWN PAUSE STOP INTENSITY INTENSITY OUTPUT 3 2 C o n n e c t i n g t h e h e a d Before turning the device on connect the head to the output socket of the device vi...

Page 8: ...e break off the therapy by pressing pause stop NOTE during the program it is possible to increase or decrease the Timer by pressing UP DOWN keys one minute each touch 3 6 1 Contact control To detect t...

Page 9: ...ery limited surface holding the handpiece firmly down on the area affected by hand or by the mechanical arm not provided together with the device which allows the operator to devote himself to other p...

Page 10: ...issue being treated 1 Bone tissue programs which deliver ultrasounds with the most common parameters for the treatment of pathologies affecting the bone structure and related structures 2 Articulation...

Page 11: ...ization 10 1 W Continue NOTE some 1 MHz therapy programs are available also with 3 MHz frequency for superficial pathologies see table except Cicatrization program wich is completely new and specific...

Page 12: ...ges it stabilizes the local metabolism and favours the afflux of the blood 10 minutes Articular stiffness Treatment of stiffness affecting the articular cartilages it restores the trophism of the cart...

Page 13: ...ndirect technique by immersion DIRECT METHOD MOBILE TECHNIQUE Area to be stimulated Notes about the position of the head Muscle Move the head all over the surface of the muscle Knee ankle Move the hea...

Page 14: ...ither supine or prone depending on the area of application This position must be kept throughout the duration of the session in order to facilitate the effects produced by the vibrations especially th...

Page 15: ...ffuse compact cellulitis Lymphatic draining 7 Localized lymphatic draining 8 Diffuse lymphatic draining Thinning 9 Local thinning 10 Diffuse thinning Skin pathologies 11 Boils 12 Granulomas 13 Local d...

Page 16: ...irculation and lymphatic draining 25 minutes Diffuse compact cellulitis Treatment of diffuse cellulitis it breaks down the fat frees the fluids which are being withheld and facilitates local circulati...

Page 17: ...utei Move the head over the whole area affected by cellulitis Thighs inside outside Move the head over the side affected by cellulitis DIRECT METHOD FIXED TECHNIQUE Area to be stimulated Notes about t...

Page 18: ...ions especially the flow of blood increased due to the vascular expansions as a consequence of the heating induced in the affected area thus facilitating the draining of the freed substances fats exce...

Page 19: ...ce to constantly check the state of wear of the insulation cable before connecting it to the power mains If it has been damaged even in part replace the cable at once Immediate maintenance New Age Ita...

Page 20: ...to the company service laboratories or else delivered to the dealer it was purchased from Service Centre New Age Italia srl Via Keplero 7 48022 Lugo RA Tel 39 0545 32019 Telefax 39 0545 369028 Website...

Page 21: ...surgical device in order to prevent danger both to the patient and to the device The instrument will work according to specification if the environment is kept at a temperature between 5 and 40 C and...

Page 22: ...h mm Weight 1 Kg Class II Type BF Classification in terms of fluid entrance IP20 Safe when there are flammable anaesthetic gases present it does not belong to categories AP or APG Appliance designed t...

Page 23: ...or to replace consumed or finished material you can purchase accessories separately 1 1 1 B a s i c B I O S O N Y C e q u i p m e n t BIOSONYC comes complete with Device 1 3 MHz 45 mm head Gel for ul...

Page 24: ...der Windt A de Winter Physiotherapy for patients with soft tissue shoulder disorders a systematic review of randomised clinical trials BMJ 1997 315 25 30 5 July A Vasta Manuale pratico illustrato di...

Reviews: