New Age BIOLASER Manual Download Page 12

Rev. 02 del 24/07/13 

12/14 

 

Keep away from children. 

 

Don't direct the handle toward the eyes

 

 

 

Ca p.9

 

–  

T E CHN ICA L  CHA R A CT E RI STI CS

 

 

–   S u p p l y    

 

Power supply: electric power PRI: 230V

 50Hz. 

Absorption from electrical power: max 60 VA. 

 

–   O u t p u t   c h a r a c t e r i s t i c s  

 

Source   : 9 diod AlGaAs 
Laser class: IV

 

Max intensity of power (P ): 4 ir x 225mW  + 5 r x 235mW 
Wavelength  (λ) :   Led infrared  λ = 940 nm  ;  Led Red  λ = 640nm 
Frequency: 100Hz 
Output  : Continuous  ( 1/1) o  modulated ON/OFF  (1/2, 1/4 ) 

 

–   O t h e r   c h a r a c t e r i s t i c s  

 

 

Dimensions: 225x170x120h [mm] 

 

Weight: 1000 [g] approx. 

 

Class: I Tipo: BF 

 

Class: II Tipo: BF 

 

Classification in comparison to the entry of liquids: IP20 

 

Safety  in presence of inflammable anaesthetic gas: it is not of category AP or 
APG 

 

Device for operation: continuous 

 

Built according to the norms:   

 

 

EN 60601-1 (1998) – Elettromedical devices : General norms for the safety 

 

EN 60601-1-2 (2003) – Collateral norm: Electromagnetic compatibility - 
Prescriptions and tests 

 

EN 60601-1-4 (1997) – Collateral norm: programmable elettromedical systems 

 

EN 980-2003 ed EN 1041 – Symbology for elettromedical instruments 

 
 
 
 
 

 
 

 
 

 
 
 
 

 
 
 
 

Serial Number (S.N)  
Identifying code for Biolaser 

 

S.N.         BLS – ZZ   XXXXX 
 

 

Identifying number of the device 

 

Year of production 

 
 

Summary of Contents for BIOLASER

Page 1: ...545 32019 Telefax 39 0545 369028 Web www newageitalia it E mail info newageitalia it This document is the property of New Age Italia All rights are reserved This document may not be copied or reproduc...

Page 2: ...ns 5 Side Effects 5 Cap 3 Operation 6 Connection to the electric net 6 Connection of the handle 7 Application of the handle 7 Switch on of the device 7 Selection of the program 7 Intensity of emission...

Page 3: ...3 Immediate maintenance 10 Cap 7 11 Cap 8 CAUTION 11 Cap 9 TECHNICAL CHARACTERISTICS 12 Supply 12 Output characteristics 12 Other characteristics 12 Cap 10 SYMBOLS 13 Cap 11 STANDARD ENDOWMENT AND AC...

Page 4: ...rough the use of the presetted programs contained in its memory and the 20 free programs Thanks to BIOLASER therefore it is possible to apply the infrared therapy with standard programs and to have th...

Page 5: ...ects with pace maker neoplasie tumori pregnancy in progress epilepsy photosensitivity children areas of growth S i d e E f f e c t s Cutaneous burns if the intensity is notable and if are present diso...

Page 6: ...modulation 6 ON OFF 7 Cable output 8 UP DOWN program selection timer adjustment C o n n e c t i o n t o t h e e l e c t r i c n e t BIOLASER can be used with battery or fed to electric current connec...

Page 7: ...he program with the keys UP DOWN see section Programs and confirm with START The programs are listed in the charts in Preset protocols I n t e n s i t y o f e m i s s i o n The intensity of the emissi...

Page 8: ...IOLASER programs table N Prog Name on display Pathology to be treated P 1 Cervical Cervical P 2 Pubalgy Pubalgy P 3 Dorsalgia Dorsalgia Bone pain P 4 Lumbago Lumbago P 5 Epicondilitis Epicondilitis P...

Page 9: ...am Repeat the operation so that to apply the therapy on the whole interested surface 4 2 S e s s i o n s Apply the laser therapy 3 4 times a week at alternate days up to the disappearance or appreciab...

Page 10: ...example if left next to intense heat sources you doubt that some liquids can be penetrated inside the wrap or other parts of the instrument are damaged broken or lacking the functionality of the instr...

Page 11: ...nd with a surgical instrument HF to avoid dangers for the patient and for the same instrument Don t use in presence of equipments for the monitoring of vital parameters The tool works according to its...

Page 12: ...c s Dimensions 225x170x120h mm Weight 1000 g approx Class I Tipo BF Class II Tipo BF Classification in comparison to the entry of liquids IP20 Safety in presence of inflammable anaesthetic gas it is n...

Page 13: ...DEVICE IS MARKED AT THE SENSES OF THE 0476 DIRECTIVE CEE 93 42 Cap 11 STANDARD ENDOWMENT AND ACCESSORIES B a s i c e n d o w m e n t BIOLASER is complete with Device Handle at 9 diods Protective goggl...

Page 14: ...d Minerva Medica Torino 1993 B Gialanella G D alessandro R Santoro Terapia fisica pratica ED Marrapese Roma 1997 A Vasta Manuale pratico illustrato di terapia fisica ED Marrapese Roma 1998 C Cisari G...

Reviews: