17
PROGR
AM INF
ORMA
TION
Program
No.
1
2
3
4
5
6
(Note 1)
7
(Note 2)
8
(Note3)
9
Frequency
(Hz)
50
50
50
35
10
ch1: 50
ch2: 10
35
8
4 – 99
Relaxation
(sec.)
5
10
20
5
5
5
5
5
Contraction
(sec.)
5
5
10
5
5
5
5
5
Length of
pulse (
µ
sec)
300
300
400
300
300
300
350
80
150
Treat Time
(mins)
30
30
30
30
30
30
30
30
Open
Indications
Strengthening Rehab
Atrophy Rehab
Sports Strengthening
Neurological Rehab
Early Muscle Activation or Detoning
Tone / Detone
Oedema Management
Facial Palsy
Acute Pain Management
Burst or
Trigger
Trigger
Trigger
Trigger
Trigger
Trigger
None
None
Trigger
Trigger
9.0
Program Information
WARNING:
The selection and setting of the program should only be made by the treating clinician.
Note 1:
For this program, the output signal sequence is as follows: Channel 1 enters a contraction cycle at frequency of 50Hz for 5
seconds and Channel 2 is off; Channel 1 is off and Channel 2 enters a contraction cycle at frequency of 10Hz for 5 seconds; Both
channels are off for a relaxation cycle of 5 seconds.
Note 2:
For this program, the output signal sequence is as follows: Channel 1 enters a contraction cycle for 5 seconds and Channel 2 is
off; Both channels are off for a relaxation cycle of 5 seconds; Channel 1 is off and Channel 2 enters a contraction cycle for 5 seconds;
Both channels are off for a relaxation cycle of 5 seconds.
Note 3:
For this program, when the amplitude is at maximum, the value written to the R2R resistor network is less than half the
maximum of 255.
The
Trigger mode
( ) is possible in Programs 1 – 5, 8 and 9. When the button is pressed the trigger mode is activated and the unit
enters a contraction cycle for as long as the key is pressed. When the key is released the unit enters the relaxation cycle. To return
to the programmed contraction/ relaxation cycle, press any of the intensity keys. The stimulation builds over a 2 second period to
the previously set intensity level.
Continuous stimulation
Ramp Down
(sec.)
1
1
1.5
1
0.5
0.5
0.5
0.5
Ramp Up
(sec.)
1
1
1.5
1
1
1
1
1
English
MediStim XP IM W Europe_PolyStim XP GA 290905.qxd 01/12/2010 12:49 Page 19
Summary of Contents for MediStim XP
Page 1: ...MediStim XP IM W Europe_PolyStim XP GA 290905 qxd 01 12 2010 12 49 Page 3 ...
Page 114: ...MediStim XP IM W Europe_PolyStim XP GA 290905 qxd 01 12 2010 12 50 Page 116 ...
Page 115: ...MediStim XP IM W Europe_PolyStim XP GA 290905 qxd 01 12 2010 12 50 Page 117 ...
Page 116: ...MediStim XP IM W Europe_PolyStim XP GA 290905 qxd 01 12 2010 12 50 Page 2 ...