Output RMS Current (RMSA):
Stands for the effective current output, which is the root mean square
current measured at a specified resistance.
Output RMS Voltage (RMSV):
Stands for the effective voltage output, which is the root mean square
voltage measured at a specified resistance.
Power (P):
Maximum power output measured in Watts (W) into a 500
Ω
load.
Frequency (F):
Number of pulses output by the unit per second, measured in Hertz (Hz).
This icon means “Warning, read the accompanying documentation”.
This symbol means “Type BF equipment”
SN stands for “serial number”.
On the rear of each XP model is the unit’s individual serial number. The letter preceding the serial number indi-
cates the year of manufacture, where “K” denotes 2005, “L” denotes 2006, etc.
This icon on your XP model shows that the device meets the 93/42/EEC Directive for medical
devices. 0366 is the number of the notified body (VDE).
!
0366
Nominal output voltage / power
Parameter
500
Ω
1k
Ω
1.5k
Ω
Output RMS voltage (RMSV)
7.5 V
12.32 V
13.74 V
Output RMS current (RMSA)
15 mA
12.3 mA
9.16 mA
Output frequency
4-99 Hz
4-99 Hz
4-99 Hz
DC Component
0 C
0 C
0 C
Pulse Width
80 – 400 µs
80 – 400 µs
80 – 400 µs
Current Intensity Range (per pulse)
0 – 75 mA
0 – 75 mA
0 – 75 mA