16. The AvivaStim XP has a load sense function that monitors the connection
between the cable/electrode and the user. When poor skin contact is detected:
• The amplitude bar of the channel being used will flash.
• The warning symbol ( ) will appear flashing on the display (Display 11).
• An audible beep will emit from the unit.
• The treatment session timer pauses.
• The intensity value falls to zero and the up-arrow intensity button is deactivated.
When proper contact is restored, stimulation builds over a 2 second period to the
previously set intensity level.
17. When the treatment is complete, the stimulation will stop automatically. You will
hear a 10 second beep alerting you that the treatment session is complete and the
display screen will appear as in Display 12. At this stage the unit should be
switched off and all electrodes removed from the body. Repackage the electrodes
carefully. The protective sheets should be stuck over the adhesive electrodes again.
Note:
The unit power will turn off automatically after 10 seconds.