Tablet Instructions
32
RNS
®
System Programming Manual
After interrogation, keep the wand in place and select
L
IVE
EC
O
G
S
from the
H
OME
screen to open the
L
IVE
EC
O
G
S
screen. An ECoG is live when you open it. Buttons provide various options for viewing the
live ECoG. Select
S
TOP
/S
TORE
to save the ECoG in the patient’s ECoG library. It remains on screen after
you stop and store it until you start another live ECoG. Captured live ECoGs are stored in the
EC
O
G
L
IBRARY
and classified as
User saved
.
Figure 21 Live ECoGs screen
Live ECoG capture can be useful for immediate or later analysis of patient ECoGs, for adjusting amplifier
gain settings, and for simulated detection as part of
Configure Pattern Detection: Define and Modify
(see page
Important Note: The recording montage must be programmed before live ECoGs can be viewed.
See Create the Recording Montage on page 32.
Note: The maximum length for a live ECoG is four minutes. If the live ECoG is allowed to scroll
for more than four minutes, only the last four minutes are stored.
Note: Live ECoGs are stored on the tablet, not in the neurostimulator memory, and are uploaded
to the PDMS if and when the tablet is logged into the PDMS.
Caution:
Telemetry Artifact
Telemetry may produce an electrographic artifact. If responsive therapy is enabled
with a sensitive detection set, detection of the electrographic artifact may occur,
resulting in therapy delivery. The physician should be aware of possible sensing
artifacts when assessing the ECoG recordings.
C
REATE THE
R
ECORDING
M
ONTAGE
The RNS
®
Neurostimulator can sense brain activity (ECoGs) from up to four amplifier channels. The
programmed configuration of leads and electrodes you assign to each channel constitute the recording
montage. You create the montage using the tablet
. There can be only one current montage and