For more clarification refer to the RCF chart below: 

RCF CHART (FIXED) 

RPM 

ROUND ROTOR 

PCR STRIP 

0.2ml 

0.5ml 

1.5 ml 

2ml 

0.2ml 

3000 

319 

424 

468 

488 

467 

4500 

718 

953 

1053 

1098 

1050 

6000 

1276 

1695 

1872 

1952 

1868 

8.  ROTOR BALANCING 

1.  Always balance the rotor before centrifugation. The following are symmetrical

loadings of centrifuge tu bes to the rotor. 

2.  Ensure thatthe samples in the tubes are of equal volume. 

3.  Symmetrical loading of tu bes in the rotor is necessary, otherwise vibration can 

occurwhich can cause serious damage to the centrifuge. 

4.  The following are the incorrect methods of loading tu bes in the centrifuge the

rotor.

5.  The following are the incorrect methods of loading tu bes in the centrifuge the

rotor.

9.  INTERCHANGEABLE ROTOR 

8 Slot Magnetic Rotor 
8 X 0.2,0.4,0.5,1.5,2 ml 

10.  CHANGING TH E ROTOR 

This centrifuge feature magnetic rotors. 

Thus changing the rotors can be done 

without any tool.  The rotors fit an the 

central  shaft  of  the  motor  through 

magnet situated at the both the rotors 

(MCT  &  PCR  rotor).  The  rotor  fits 

through  snap  to  the  central  shafts 

through  gentle  guidance  into  the 

groove  an the rotor  adapter  (Pull  the 

rotor  upwards with both hands  gently 

and replace it with required rotor) . 

• 

PCR Magnetic Rotor 

16x0.2ml 

Summary of Contents for iFUGE M08VT

Page 1: ...NEUATION TECHNOLOGIES j iFUGE MOSVT PRODUCT USER MANUAL ...

Page 2: ...hould not attempt to service or repair the unit ffrepairsarerequired pleasecontactthesupplier 3 FEATURES 8positionmicrocentrifugetuberotor Digitaldisplaywithset variablespeed timer Electronicsafetybrakeonlid openingforusersafety Magneticrotorfor easytool freechangingofrotor Startsand stopsinseconds Lidsafetyinterlock Smoothandsilent discreetrunning Smallfootprintcanbe fitinhoods smallworkplaces In...

Page 3: ...ge whileit sunderoperation Do not fill the tubes while they are in the rotor Liquid spillage may harm the unit Do not put hands in the rotor chamber unless the rotor has completely stopped Donotcentrifuge flammable explosiveor corrosive materials Do not use solvents or flammable media near this or other electrical equipment In case the unit is dropped from the table please check the casing for any...

Page 4: ...trifuge 4 The following are the incorrect methodsof loading tubes in the centrifuge the rotor 5 The following are the incorrect methodsof loading tubes in the centrifuge the rotor 9 INTERCHANGEABLE ROTOR 8 Slot Magnetic Rotor 8 X0 2 0 4 0 5 1 5 2ml 10 CHANGINGTHE ROTOR This centrifuge feature magnetic rotors Thus changing the rotors can be done without any tool The rotors fit an the central shaft ...

Page 5: ... defects in original material or workmanship This warranty does not cover any incidental or consequential damages commerciallossor any other damages fromthe useofthisproduct The warranty is invalidated by any non factory modification which will immediatelyterminate all liabilities on us for the productsor damages caused by its use The customer shall be responsible for the product or use of product...

Page 6: ...ations based on EU Directive 2012 19 EU on waste electrical and electronic equipment WEEE According to these regulations any devices supplied after 13 06 05 in the business to business sphere to which this product is assigned may no longer be disposed off in municipal or domestic waste They are marked withthe followingsymboltoindicatethis As disposal regulations within the EU may vary from country...

Reviews: