6. SAFETY PRECAUTIONS
The personal micro centrifuge is designed to be safe under the following
conditions when operated with necessary precautions mentioned below:
• Indoor use only.
• Ambient temperature of4'Cto45'C.
• Main supply voltage fluctuations not to exceed +/- 10% of the nominal
voltage.
• Position at least 20 cm from walls, 20 cm from ceiling and 20 cm between
other products.
• Make sure the rotor is placed in the shaft groove properly. If the rotor is not
installed properly, the machine will make noise, turn off the machine &
install the rotor as instructed for fitment of the rotor.
• Always load the rotor symmetrically. Each tube should be counterbalanced
by another tube.
• Make sure to always use only specified micro centrifuge tubes made from
plastic and designed to withstand centrifugal forces of at least 2000 g.
• Do not operate the centrifuge without the rotor being properly attached to
the shaft. Re-check by trying to pull it up manually, especially after the rotor
has been changed or re-assembled.
• Do not move the centrifuge while it's under operation.
• Do not fill the tu bes while they are in the rotor. Liquid spillage may harm the
unit.
• Do not put hands in the rotor chamber unless the rotor has completely
stopped.
• Do not centrifuge flammable, explosive or corrosive materials.
• Do not use solvents or flammable media near this or other electrical
equipment.
• In case the unit is dropped from the table, please check the casing for any
cracks/damage.
• The centrifuge and its components are non autocleavable.
• Do not use the centrifuge in any manner not specified in these instructions.
• Use mild soap solution to clean & disinfect. Do not using sanitizer / !PA
solutions.
7. INSTALLATION
ON/OFF
....-�......,.-',--
SWITCH
ROTOR
---.+++.-
POWER
This micro centrifuge is supplied in a packaging box. Open the box, then gently
remove the centrifuge from the box and plug in the power adaptor pin before
connecting to the main power source. The instruction manual and the
accessories should be kept with the centrifuge. Please keep all packaging in safe
storage for at least 2 years for warranty purposes.
Make sure the rotor is tight to avoid any accident. Load the rotor with the micro
tu bes for centrifugation. Place the ON/OFF switch to "o" on back side and close
the lid to start centrifugation .
•