TalT:

 

Tall T Wave Rejection.

 QRS and T waves are generated with the following values: 

 
QRS: Rate  

 80 BPM 

Amplitude 

 1 mV 

 Width    

 100 ms 

 

T Wave Duration   180 ms 

 

QT Interval 

 350 ms 

T Wave Amplitude 0.0, 0.2, 0.4, 0.6, 0.8, 1.0,  and 1.2 mV .   
 
In the automatic test sequence the T Wave Amplitude steps through the changes at one minute 
intervals. The display indicates the T Wave Amplitude and the QRS Rate. 
 
As the T wave amplitude increases, the first value at which the patient monitor counts the T wave at 80 

±

 8 BPM should be noted.  This value should match the patient monitor manufacturer’s specification. 

 

PPR: Pacemaker Pulse Rejection.

  The test sequence cycles through normal paced rhythm (AP), 

ineffective pacing (Asynchronous Pacing ASP), and pacemaker pulses alone (PCR).  
 
For normal pacing (AP),QRS and pacer waves are generated with the following values: 

 

 

QRS: Amplitude       1 mV 

 

          Width  

     100 ms 

 

T wave: Amplitude     0.2 mV 

 

     

              Duration       180 ms 

 

Q-T Interval 

     350 ms 

 

R-R Interval 

     1 Sec 

 

Pacer: Amplitude       2 mV, -2 mV 

 

            Width 

     2 ms, 0.1 ms 

 
For ineffective pacing (ASP), the values of the QRS and pacer waves are the same as normal pacing 
except for the QRS rate that becomes 30 BPM and the pacer rate that becomes 80 BPM.   
 
During the ASP and AP test sequences the display will show the pacer amplitude, the pacer width, and 
the QRS rate. 
 
For pacemaker pulses alone (PCR) the values generated are: 
 

    Pacer: Rate  

     60 BPM 

 

 

 Width       2.0 ms, 0.1 ms 

 

 

 Amplitude  2 mV, -2 mV   

 
During the PCR test sequence the display will show the pacer amplitude, the pacer width, and the QRS 
rate.  
Each set of values in the test sequence is generated and displayed for 20 seconds. 
 

TAC:

 

Time to Alarm for Tachycardia.

 The TAC test is designed to measure the time it takes for the 

patient monitor to alarm after the onset of ventricular tachycardia. The low and high alarms on the 
patient monitor should be set at 60 BPM 100 BPM before starting the test. 
 
This auto test sequence generates a QRS wave form at the rate of 80 BPM alternating with a ventricular 
tachycardia waveform with rates of 206 and 195 BPM and amplitudes of 1.0, 0.5, 2.0, and 4.0 mV. 
 
Each waveform is generated for 20 seconds and the display will show the amplitude and the rate of the 
waveform being generated. 
 
 
The following is the sequence of waveforms: 
       QRS 

Ventricular Tachycardia 

 Rate 

Amp 

Width 

Rate  

Amp 

(BPM) (mV)   

(ms)  (BPM) (mV) 

  80 

   1 

  100 

  206 

 1.0 

  80 

   1 

  100 

  206 

 0.5 

  80 

   1 

  100 

  206 

 2.0 

Minisim 1000 User Manual

 15

 

 

Summary of Contents for MiniSim 1000

Page 1: ...MiniSim 1000 Multi Parameter Patient Simulator INSTRUCTION MANUAL 110 Toledo Street Farmingdale N Y 11735 Phone 631 531 0100 Fax 631 531 0101 www Netech org Minisim user manual...

Page 2: ...a product that proves to be defective provided you return the product shipping prepaid to Netech Corporation This warranty does not apply if the product has been damaged by accident or misuse or as th...

Page 3: ...rademarks of Netech Corporation Any other trademark names used in this manual are only for editorial purposes and the benefit of the respective trademark owner with no intention of improperly using th...

Page 4: ...ies 10 2 Operating Instructions Controls and Indicators 08 Baseline ECG Waveform 09 Performance Waveforms 11 Arrhythmia Waveforms 12 Automatic Testing Sequences 13 Respiration Waveforms Apnea 16 Blood...

Page 5: ...libration traceable to the NIST The MiniSim 1000 is a rugged instrument that performs its simulations quickly accurately and with ease 1 2 Specifications ECG 12 leads with independent outputs referenc...

Page 6: ...60 and 120 BPM Accuracy 1 Apnea Off continuous 12 seconds and 32 seconds Lead Configuration Leads I and II BLOOD PRESSURE Impedance 350 Ohms Excitation 2 to 16 Volts Sensitivity 5 V V mmHg Static 0 5...

Page 7: ...3 ACCESSORIES Description Part Number User Manual 300 Manual Pressure Cable Open Ended 303 Temperature Cable Open Ended 304 OPTIONAL ACCESSORIES Description Part Number Soft Carrying Case 301 110 VAC...

Page 8: ...rature Default Setting YSI 700 series set to 25 C YSI 400 series set to 25 C CONTROLS AND INDICATORS 1 On Off Key 2 ECG Lead Snaps 3 LCD Display 4 Temperature Connector 5 Blood Pressure Connector 6 Hi...

Page 9: ...ameter for simulation ECG Resp BP or Temp 4 Continue to select choices under the parameter to be simulated until the functional characteristics have been chosen 5 Each Parameter has default settings E...

Page 10: ...with pacer waves with a 0 1 ms or 2 ms width and amplitude of 2 mV at 80 BPM Note 0 1 ms pacer width for AP and ASP can only be selected by going to PCR A selection other than 0 1 ms produces 2 ms wi...

Page 11: ...orm with the Q wave modified as in Infarction and the ST segment elevated 7 as in ST Elevation Selecting TalT branches to TalT Rate 80 Rate Ampl An 80 BPM QRS test signal of 1 mV amplitude and 100 ms...

Page 12: ...at Normal Sinus Rhythm is generated at 80 BPM with every 10th beat missing AFLT Atrial Flutter Varying ECG rates with 12 cycles at 60 BPM for 12 seconds 9 cycles at 90 BPM for 6 seconds 15 cycles at 1...

Page 13: ...R is generated at 80 BPM Each time the F4 function key is pressed 12 PVCs are generated at different intervals PV24 Premature Ventricular Contraction 24 NSR is generated at 80 BPM Each time the F1 fun...

Page 14: ...equence Amplitude Width Rate Time mV ms BPM sec 0 5 100 80 30 210 20 70 80 30 210 20 120 80 30 210 20 2 0 100 80 30 210 20 70 80 30 210 20 120 80 30 210 20 5 0 100 80 30 210 20 70 80 30 210 20 120 80...

Page 15: ...cept for the QRS rate that becomes 30 BPM and the pacer rate that becomes 80 BPM During the ASP and AP test sequences the display will show the pacer amplitude the pacer width and the QRS rate For pac...

Page 16: ...ormal respiration waveforms are generated Cont Continuous Apnea No respiration waveforms are generated 12s No respiration waveform is generated for 12 seconds 32s No respiration waveform is generated...

Page 17: ...nd YSI 700 CALIBRATION The MiniSim 1000 is shipped from the factory fully calibrated with a Certificate of Calibration traceable to the NIST Annual calibration is recommended Contact the Customer Serv...

Page 18: ...Appendix A Minisim 1000 User Manual 18...

Reviews: