LKG 601
Netech Corporation
110 Toledo St, Farmingdale, NY 11735,
Specifications
Display
: 3-1/2 Digit LCD Display
LEAKAGE CURRENT:
0-1999 Micro Amps. All Current measurements are
made through the AAMI/ANSI ES1-1985 or IEC 601-1 test load. The meter
readings correspond to the true RMS value of the current.
Accuracy:
± 1% of reading, + 1 LSD: DC to 1 KHz,
± 2.5% of r 1 LSD
1 KHz to 100 KHz,
± 5% of r 1 LSD
100 KHz to 1 MHz
RESISTANCE:
0 TO 1999 Milli Ohms.
Accuracy:
+ 1 % of r 1 LSD
DEVICE CURRENT:
0 to 19.99 Amps
Accuracy:
+ 2% Full Scale + 1 LSD
TEST RECEPTACLE:
Hospital Grade 110 VAC -15 Amp or 220 VAC 10 Amp.
The rocker polarity switch selects NORMAL, OFF and REVERSE polarity to the
test receptacle, and a momentary Neutral switch will open the neutral line to the test
receptacle.
POWER REQUIREMENTS:
110 VAC 50-60Hz, 15 Amps or 220 VAC, 10 Amps
maximum rated for the test outlet and 0.1 Amp rated for the unit.
PHYSICAL DIMENSIONS:
Size: 5.5x 3.25 x 2.5 Inches
Weight: 1 lbs (.45kg)
ENVIRONMENTAL:
Operating range: 59 to 950 F (15 to 400 C) Storage
Temperature: 0 to 1220 F (-20 to 600 C) Relative Humidity: 90% (max) at
temperatures