background image

 

 

LKG 601 

  

 

 

 

 

 

Netech Corporation 

110 Toledo St, Farmingdale, NY 11735,  

 

http://www.NetechCorp.us

 

http://www.PressureMeter.com

 

 

 

Specifications 

 

Display

: 3-1/2 Digit LCD Display 

 

      LEAKAGE  CURRENT: 

0-1999  Micro  Amps. All  Current  measurements  are 

made  through  the AAMI/ANSI  ES1-1985  or  IEC  601-1  test  load. The  meter 

readings correspond to the true RMS value of the current. 

 

 

Accuracy: 

± 1% of reading, + 1 LSD: DC to 1 KHz, 

± 2.5% of r 1 LSD 

1 KHz to 100 KHz, 

± 5% of r 1 LSD 

100 KHz to 1 MHz 

 

RESISTANCE: 

0 TO 1999 Milli Ohms. 

Accuracy: 

+ 1 % of r 1 LSD 

 

DEVICE CURRENT: 

0 to 19.99 Amps 

Accuracy: 

+ 2% Full Scale + 1 LSD 

 

TEST RECEPTACLE: 

Hospital Grade 110 VAC -15 Amp or 220 VAC 10 Amp. 

 

The rocker polarity switch selects NORMAL, OFF and REVERSE polarity to the 

test receptacle, and a momentary Neutral switch will open the neutral line to the test 

receptacle. 

 

POWER REQUIREMENTS: 

110 VAC 50-60Hz, 15 Amps or 220 VAC, 10 Amps 

maximum rated for the test outlet and 0.1 Amp rated for the unit. 

 

 

PHYSICAL DIMENSIONS:

 Size: 5.5x 3.25 x 2.5 Inches 

                                                        Weight: 1 lbs (.45kg) 

 

ENVIRONMENTAL:

  Operating range: 59 to 950 F (15 to 400 C) Storage 

Temperature: 0 to 1220 F (-20 to 600 C) Relative Humidity: 90% (max) at 

temperatures 

Summary of Contents for 510-110

Page 1: ...LKG 601 Netech Corporation 110 Toledo St Farmingdale NY 11735 http www NetechCorp us http www PressureMeter com LKG 601 ELECTRICAL SAFETY ANALYZER INSTRUCTION MANUAL...

Page 2: ...pair and Recalibration 5 Models and Part Numbers 6 General Overview Introduction 8 Key Features 8 Specifications 9 Instrument Familiarity 10 Controls Indicators 11 Preparation for use 12 Operating Ins...

Page 3: ...otice and without incurring any obligation Netech guarantees availability of service parts for 5 years after the manufacture of the unit is discontinued The warranty is void if you elect to have the u...

Page 4: ...st procedures are found to meet those specifications when it was shipped from the factory Calibration LKG 601 is calibrated using standards traceable to National Institute of Standards and Technology...

Page 5: ...return or refund Only products returned within 60 days from the date of original purchase are eligible for refund credit In order to receive a partial refund credit of a product purchase price on a s...

Page 6: ...aging or handling Use the original carton and packaging material for shipment Returns for Refund Credit A Return Material Authorization RMA number must be obtained from our service or customer service...

Page 7: ...Hard case 510 220 Electrical Safety Analyzer 220 V with Hard case 510 110 220 Electrical Safety Analyzer 110 V and 220V Operation with Hard Case Standard Accessories Item Description Picture 503 Test...

Page 8: ...ects the test to be performed The user may select either the AAMI ESI 1993 or the IEC 601 1 test load to compensate for high frequency components in the measurement Accurate resistance measurements ar...

Page 9: ...TO 1999 Milli Ohms Accuracy 1 of reading 1 LSD DEVICE CURRENT 0 to 19 99 Amps Accuracy 2 Full Scale 1 LSD TEST RECEPTACLE Hospital Grade 110 VAC 15 Amp or 220 VAC 10 Amp The rocker polarity switch se...

Page 10: ...User Manual LKG 601 10 Instrument Familiarity 1 3 6 5 2 4 7 9 8 13 11 10 12...

Page 11: ...pen the neutral line to the test receptacle 7 Load Selector Switch A slide switch on the right side of the instrument to select either the AAMI or the IEC 601 1 test load 8 Test Receptacle A 20 AMP po...

Page 12: ...n of the dual voltage LKG 601 Before proceeding with any measurement become familiar with the measurements and the function selector switch During the initial setup the last digit of the LCD will chan...

Page 13: ...to CORD RESISTANCE Plug the Device Under Test into the test outlet Using the test lead provided connect the chassis of the Device Under Test to the Case Ground jack on the LKG 601 The display will in...

Page 14: ...e case of the DUT Device under test The leakage measurement is made when the Ground switch is momentarily opened The LCD will display the measurement in Micro Amps Both Earth and Case Leakage Current...

Page 15: ...indicating that the cord circuit is open 2 Move the Mode selector switch to all three positions and test the LED indication 3 Check the continuity between receptacle ground and case ground using the...

Page 16: ...switch selects leakage current measurement the AAMI LOAD is connected in front of the INPUT AMPLIFIER and the voltage developed across it from the leakage current is measured Current Source and Amplif...

Page 17: ...libration is recommended If the LKG 601 is returned to Netech for recalibration before the first year warranty expires Netech will provide a second year warranty The warranty is void if the LKG 601 is...

Page 18: ...A number follow the link http www NetechCorp us and fill in the required information Email service netechcorp us or call 800 547 6557 US Canada International 631 531 0100 When shipping units to the fa...

Page 19: ...dapter for 220 Volt devices CAUTION LKG 601 dual voltage model PN 510 110 220 designed for use in 110 Volt Ac powered equipments or 220 AC powered equipments For 220 volt devices make sure the device...

Page 20: ...User Manual LKG 601 20 Appendix This section is intentionally left blank for future use...