LKG 601
Netech Corporation
110 Toledo St, Farmingdale, NY 11735,
Operating Instructions
Cord Resistance Measurement:
See Figure 2.
•
Set the polarity switch to the off (center) position.
•
Move the MODE switch to CORD RESISTANCE. Plug the Device Under
Test into the test outlet.
•
Using the test lead provided, connect the chassis of the Device Under Test
to the Case Ground jack on the LKG 601.
•
The display will indicate the Power Cord Resistance in milliohms.
Hammer pressure levels in excess of 20 PSI will cause damage t
Figure 2
Block Diagram for Cord Resistance measurement and
Case Leakage measurement (external)
Case Ground
Power Cable (DUT)
Device Under
Test