CLEANING AND MAINTENANCE
Contamination of the optical surfaces at the outlet openings, the beam quality and
reach a decisive influence. Cleaning with cotton swab and looking for lint.
CHECK THE ACCURACY OF THE PULSAR H
1. Attach the PULSAR H on a tripod or horizontal plane around 0.5 m
parallel and perpendicular from a wall A and 30 meters away from a wall B.
2. Make sure that the surface on which the laser device is located roughly aligned.
3. Turn on the PULSAR H.
4. Wait until the unit has itself leveled.
5. Mark the position of the laser beam on the wall A.
6. Now turn over the laser 180 ° and make a mark to the opposite wall B.
7. Next, position the device about 0.5 m from the wall B.
8. Now adjust the height of the laser system so that the laser beam hits the mark
exactly.
9. Now turn over the laser 180° and align the laser beam on the wall A. Mark the
position of the laser beam on wall A.
10. If the marks on the wall A match the laser device is optimally calibrated.
11. Should the markings disagree, measure the difference. To determine the accuracy
of sharing the value determined by two. The new value corresponding to a
measurement error of the distance between the two walls. If the values exceed
2 mm to 10 m, a calibration is necessary.