Product 

D-6026 

Silicone adapter   

D-6040 

Nose cone for neoPipetta 

D-6028 

Filter 0.45µm 

D-6029 

Battery 

 

The neoPipetta cannot be operated without a filter. Wetting of filter due to overfilling will slow or prevent fluid 

movement. If overfilling occurs, change the filter immediately. Only use filters supplied by the manufacturer. 

6.3. Changing the battery   

If the battery needs be replaced follow these steps to replace the battery: 

 

Take the lid off the battery compartment to expose the battery compartment   

 

Disconnect the plug of the Lithium-ion battery from the Printed Circuit Board (PCB) connector inside 

the battery compartment exposing a 2pin connector on the PCB   

 

Connect the 2 pin plug of the new Lithium-ion battery to the exposed 2 pin connector on the PCB.    The 

side of the plug with the exposed metal tabs, should face away from the PCB. If the battery is connected 

with the plug in the wrong orientation, the indication arrow light will be not on. 

 

                                       

                                           

   

 

 

 

Replace the battery compartment lid. 

 

Dispose of the battery appropriately. 

     

  Note:

 Only use the original battery supplied by the manufacturer. 

7. STORAGE 

The  neoPipetta  should  not  be  stored  for  a  long  period  of  time  without  removing  the  serological  pipet.  It  is 

recommended to disconnect the neoPipetta from the AC-adapter if the unit is not used for a lengthy period of time. 

Fig.2 

Power 

display 

 

Fig.4 

Fig.3 

Side with expose 

metal tabs 

2 Pin connector 

on PCB 

Summary of Contents for neoPipetta D-6019

Page 1: ...neoPipetta D 6019 User Manual...

Page 2: ...ne and silicone adaptor 2 4 4 Filters 2 5 OPERATING THE NEOPIPETTA 2 5 1 Setting the operating speed 2 5 2 Inserting a serological pipet 2 5 3 Aspiration 3 5 4 Dispensing 3 5 5 Other pipetting recomme...

Page 3: ......

Page 4: ...ltage of your electrical outlet Connect the AC adapter to the electrical outlet and to the electrical fitting at the base of the handle of the neoPipetta Note The neoPipetta should only be charged wit...

Page 5: ...ptor exposes a hydrophobic filter 0 45 micron located at the top of the nose piece The filter protects the neoPipetta from accidental over aspiration It is strongly recommended to use only the origina...

Page 6: ...ument in an atmosphere with a danger of explosion Never use force with the neoPipetta Should the instrument fail to function properly stop pipetting and check the Trouble Shooting section 6 MAINTENANC...

Page 7: ...the battery compartment exposing a 2pin connector on the PCB Connect the 2 pin plug of the new Lithium ion battery to the exposed 2 pin connector on the PCB The side of the plug with the exposed meta...

Page 8: ...d in the wrong orientation Failure of the Printed Circuit Board Replace battery Refer to section 6 3 of manual to plug battery in correctly Call the manufacturer Tech Support 9 WARRANTY INFORMATION Th...

Page 9: ...neoLab Migge GmbH Rischerstr 7 9 69123 Heidelberg Germany Tel 49 06221 8442 0 Fax 49 06221 8442 9933 Email info neolab de www neolab de...

Reviews: