3
5.3. Aspiration
Select a speed and fill the pipet by pressing the aspirating button. Fill the pipet just above the desired
level, check level by the meniscus.
Remove the serological pipet from the sample.
Adjust the volume of the sample by pressing the dispensing button until the meniscus is at the required
level.
5.4. Dispensing
The neoPipetta offers two different dispensing modes:
If the dispensing button is pressed only partially to a point where a check can be felt, the dispensing
function will be based on the force of gravity and the liquid will simply flow out.
Pressing the dispensing button further in will engage the motor and activates the blow-out function.
Note:
Finger pressure can be varied to fine-tune the speed within each mode.
5.5 Other pipetting recommendations
Use the instrument within the limits of its chemical resistance.
Never use with highly flammable liquids.
Observe the reagent manufacturer’s safety information.
Never use the instrument in an atmosphere with a danger of explosion.
Never use force with the neoPipetta.
Should the instrument fail to function properly, stop pipetting and check the Trouble Shooting section.
6. MAINTENANCE
The neoPipetta has been designed for easy in-house maintenance.
6.1. Decontamination
To decontaminate, lightly spray the outside of the neoPipetta with the decontamination spray and wipe dry with a
clean cloth.
6.2. Changing the nose cone, silicone adapter and filter (Fig. 2)
To change the nose cone, silicone adapter and filter firmly hold the nose cone by the ribbed collar, rotate the nose
cone counter-clockwise until completely loosen and pull out the nose cone and the silicone adapter. The filter can
be removed by simply pulling it out once the nose cone and the silicone adapter have been disengaged. Position
the new filter with the colored side upwards. To reassemble, insert the adapter and nose cone and twist the nose
cone gently clockwise.
Note:
It is recommended to check and clean the silicone adapter every six months, or when necessary, to prevent a
loss of vacuum. If there is a loss of vacuum, please contact the manufacturer or your local distributor.