Technical data: 
 

Voltage: 220-240 
Frequency Hz: 50/60 

 

 

The safety temperature range of the hotplate  100-350 [C ] 

 

Temperature sensor in medium PT1000 

 

Control accuracy of heating temperature with temperature sensor 0,2 C 

 

Dimensions (mm) 2 8 0 x 1 6 0 x 8 5 

 

Permitted ambient temp.  5-40°C 

 

Permitted relative humiditv: 80% 

 

Protection class acc. to DIN 60529: IP 42 

 

RS232 interface: Yes 

 

Stirring point position: 1 

 

Max. stirring quantity(H2O): 20 L. 

 

Max. magnetic bar: 80 mm 

 

Motor type: External rotor, brushless motor 

 

Max. power input of motor: 18 W 

 

Max. power output of motor: 10 W 

 

Speed range [rpm]: 100~1500 

 

Rotary speed display: LCD 

 

Speed display accuracy [rpm]: 1 

 

Hotplate material: stainless steal 

 

Hotplate diam.: 135 mm 

 

Heating power: 500 W 

 

Temperature range (C): RT~340 

 

Temperature display accuracy: 0,1°C 

 

Control accuracy of heating temperature: 1°C 

 
 
 
 
 
                            neoLab Migge Laborbedarf-Vertriebs GmbH, Rischerstr. 7-9, 69123 Heidelberg  
                                         Tel. 06221 / 84 42-0      [email protected]      www.neolab.de      
 
 
 
         neoLab Berlin Tel. 030 / 30 87 45-0                                     neoLab München Tel. 089 / 35 61 88 -0 
 

Summary of Contents for D-6010

Page 1: ...oof surface do not operate the instrument in explosive atmospheres with hazardous substances or under water Gradually increase the speed reduce the speed if The stirring bar breakaway because of too high speed The instrument is not running smoothly or container moves on the base plate Temperature must always be set to at least 25 C lower than the fire point of the media used Beware of hazards due ...

Page 2: ...ice and the wall and above the assembly min 100 mm Proper use The instrument is designed for mixing and or heating liquids in schools laboratories or factories This device is not suitable for using in residential areas or other constraints mentioned in Chapter 1 Inspection Unpack the equipment carefully and check for any damages which may have arisen during transport If it happens please contact n...

Page 3: ...emperature value in about 5 seconds Remote control The unit can be controlled from an external PC using the dedicated software via the RS232 C serial interface fitted to the unit Data communication from laboratory instrument to computer is only possible on demand of the computer The functions of the interface lines between laboratory instrument and automation system are selected from the specified...

Page 4: ...temperature is above 50 C the LCD still work and Hot flash Turn off residual heat warning function before powered OFF the unit If these faults are not resolved please send the instrument back Maintenance and Cleaning Proper maintenance can keep instruments working in a good state and lengthen its lifetime Be careful not spray the cleanser into the instrument when cleaning Unplug the power line whe...

Page 5: ...stirring quantity H2O 20 L Max magnetic bar 80 mm Motor type External rotor brushless motor Max power input of motor 18 W Max power output of motor 10 W Speed range rpm 100 1500 Rotary speed display LCD Speed display accuracy rpm 1 Hotplate material stainless steal Hotplate diam 135 mm Heating power 500 W Temperature range C RT 340 Temperature display accuracy 0 1 C Control accuracy of heating tem...