Users should read this Manual carefully, follow the instructions and procedures, and be aware of the 
dangers while using this instrument. 

 
Safety Instructions 

Warning! 

 

 

Read the operating instructions carefully before use the instrument 

 

Ensure that only trained staff work with the instrument 

 

Forbid to heat the substances with low burning point or easy-volatile( hotplate ) 

                Risk of bum! 

 

Caution when touching the housing parts and the heating plate. 

        The heating plate can reach temperatures of 340°C! 

 

Pay attention to the residual heat after switching off. 

 

Protective ground contact! 

 

 

Make sure that socket must be earthed (protective ground contact) before use 

 

When work, wear suitable protective clothing to avoid the risk from: 

 

Splashing and evaporation of liquids 

 

Release of toxic or combustible gases. 

 

Set up the instrument in a spacious area on a solid, clean, non-slip, dry and fireproof surface, do not 
operate the instrument in explosive atmospheres with hazardous substances or under water. 

 

Gradually increase the speed, reduce the speed if : 

 

The stirring bar breakaway because of too high speed 

 

The instrument is not running smoothly, or container moves on the base plate. 

 

Temperature must always be set to at least 25°C lower than the fire point of the media used. 

 

Beware of hazards due to

 

 

Flammable materials or media with a low boiling temperature 

 

Overfilling of media 

 

Unsafe container 

 

Process pathogenic materials only in closed vessels. 

 

If the case of the stirrer bar is PTFE, please note: 

Elemental fluorine, three fluoride and alkali metals will corrode the PTFE and Halogen Alkanes make it expansion 
at room temperature. Molten alkali, alkaline earth metals or their solution, as well as the powder in second and 
third ethnic of the periodic table of Elements will have chemical reaction with PTFE when temperature reaches 
300 – 400°C. 
 
 
                            neoLab Migge Laborbedarf-Vertriebs GmbH, Rischerstr. 7-9, 69123 Heidelberg  
                                         Tel. 06221 / 84 42-0      [email protected]      www.neolab.de      
 
 
 
         neoLab Berlin Tel. 030 / 30 87 45-0                                     neoLab München Tel. 089 / 35 61 88 -0 
 

Summary of Contents for D-6010

Page 1: ...oof surface do not operate the instrument in explosive atmospheres with hazardous substances or under water Gradually increase the speed reduce the speed if The stirring bar breakaway because of too high speed The instrument is not running smoothly or container moves on the base plate Temperature must always be set to at least 25 C lower than the fire point of the media used Beware of hazards due ...

Page 2: ...ice and the wall and above the assembly min 100 mm Proper use The instrument is designed for mixing and or heating liquids in schools laboratories or factories This device is not suitable for using in residential areas or other constraints mentioned in Chapter 1 Inspection Unpack the equipment carefully and check for any damages which may have arisen during transport If it happens please contact n...

Page 3: ...emperature value in about 5 seconds Remote control The unit can be controlled from an external PC using the dedicated software via the RS232 C serial interface fitted to the unit Data communication from laboratory instrument to computer is only possible on demand of the computer The functions of the interface lines between laboratory instrument and automation system are selected from the specified...

Page 4: ...temperature is above 50 C the LCD still work and Hot flash Turn off residual heat warning function before powered OFF the unit If these faults are not resolved please send the instrument back Maintenance and Cleaning Proper maintenance can keep instruments working in a good state and lengthen its lifetime Be careful not spray the cleanser into the instrument when cleaning Unplug the power line whe...

Page 5: ...stirring quantity H2O 20 L Max magnetic bar 80 mm Motor type External rotor brushless motor Max power input of motor 18 W Max power output of motor 10 W Speed range rpm 100 1500 Rotary speed display LCD Speed display accuracy rpm 1 Hotplate material stainless steal Hotplate diam 135 mm Heating power 500 W Temperature range C RT 340 Temperature display accuracy 0 1 C Control accuracy of heating tem...