TROUBLESHOOTING

 

P

ROBLEM

 

TRY THIS

 

Vitality user interface screen is not 
activated when the Smartcable is plugged 
into phone. 

Delete and re-install the OHA app. Occasionally 
updates in other apps that access the charging 
port can interfere with the phone properly 
recognizing the Smartcable. 

The Vitality screen has been replaced 
with other screens from the OHA app.  

1.

Swipe the OHA screen to the left and the Vitality

 

screen should appear.

2.

If not try deleting and re-installing the OHA app.

I  cannot  feel  any  electrical  pulsations 
even after I have increased the voltage to 
100V. 

1.

Make sure the Smartcable is powered on. When

 

active, the time will be counting down and the

 

yellow and green lines will be oscillating.

2.

Be sure to wipe your ear and the contact surface

 

of the ear clip with a moist alcohol swab before

 

each 

use. The oil on these surfaces can block the

 

conduction of electricity. Some inexpensive brands

 

of alcohol swabs have little alcohol on them and do

 

not remove the oil effectively.

3.

Replace the ear clip every 3 months. Carbon-

rubber electrodes may lose their conductivity over

 

time. Discard the ear clip if its contact surface

 

becomes cracked, worn or if the electrode face loses

 

its sheen.

After plugging in the Smartcable, the 
installing arrow does not stop rotating 
and the Vitality user interface screen does 
not appear. 

1.

Make sure the Smartcable is fully inserted into

 

the phone

s charging port.

2.

Try deleting and re-installing the OHA app.

3.

Installing the OHA app and running the

 

Smartcable on another phone can help you

 

determine if the problem might be in your phone

s

 

charging port.

4.

If it still does not operate correctly, the cable

 

might be damaged.  Contact
[email protected].

The Smartcable does not fit the 
charging port on my Android phone. 

The Smartcable has a built-in micro-USB male 
plug.  Some Android phones require a USB-C male 
plug.  Use the micro-USB to USB-C adapter that 
was included with your purchase of the Vitality 
Smartcable. 

Summary of Contents for Vitality SMARTCABLE

Page 1: ...USER GUIDE Neuromodulation Product Description Getting Started Safety Troubleshooting Return Policy Warranty ...

Page 2: ...e health Electrical neuromodulation is the newest method to stimulate the mind and improve health Electric neuromodulation has been safely used in many forms for several decades as an alternate method to stimulate the mind The concha of the ear is used as a neuromodulation point in acupuncture and acupressure and has been discovered to be an excellent access point for electric neuromodulation The ...

Page 3: ...odulation The Vitality Smartcable works by sending proprietary low voltage micro pulses of electrical current to an ear clip with two opposing rubber electrodes The ear clip is designed to be clipped to the tragus or the concha of the outer ear It s necessary to use high quality alcohol wipes to clean the ear and the ear clip before each session to ensure effective contact The Vitality Smartcable ...

Page 4: ...gus or concha of the ear Attach the collar clip to take the weight of the cable off your ear 7 Tap the Start button to switch app on 8 Increase voltage until you feel a current Most users will feel some electric tingling at the ear around 30 50V Occasionally a user may need to increase the voltage to higher levels to be felt The difference in sensing the electricity is due to fluctuations in moist...

Page 5: ...ot been evaluated Store your Vitality Smartcable a safe location out of reach of children Do not place the unit close to excessive heat Contact your healthcare provider if you have any questions regarding your health or symptoms you may be experiencing Consult your healthcare provider before using on children under the age of 18 Do not use on children under the age of 5 CONTRAINDICATIONS Safety an...

Page 6: ...t only 2 6 of participants discontinued use due to side effects where the most common side effect reported was skin irritation at the stimulation site which typically ceased shortly after use The following side effects were reported in this review Tingling pain redness itching at the site of electrode placement during long term application 16 7 Lightheadedness dizziness 1 4 Fatigue 0 5 Shortness o...

Page 7: ...icity Some inexpensive brands of alcohol swabs have little alcohol on them and do not remove the oil effectively 3 Replace the ear clip every 3 months Carbon rubber electrodes may lose their conductivity over time Discard the ear clip if its contact surface becomes cracked worn or if the electrode face loses its sheen After plugging in the Smartcable the installing arrow does not stop rotating and...

Page 8: ...to the returned item s while in transit back to Nemechek Technologies Vitality Smartcable ONE YEAR LIMITED WARRANTY Last Updated August 2019 Nemechek Technologies LLC Nemechek Technologies warrants to the original purchaser that the Vitality Smartcable Stimulator device the Product will be free from defects in materials and workmanship under normal use for one 1 year from the date of original purc...

Page 9: ...n the Warranty Period at Nemechek Technologies s option a repair the Product at no charge using new or refurbished replacement parts b replace the Product with a new or refurbished Product or c refund the customer the purchase price paid for the Product if Nemechek Technologies determines that it is not reasonable to repair or replace the defective Product Except where prohibited by applicable law...

Page 10: ...solution provider mutually agreed upon by the parties The alternative dispute resolution provider and the parties must comply with the following rules a the arbitration shall be conducted by telephone online and or be solely based on written submissions the specific manner of which shall be chosen by the party initiating the arbitration b the arbitration shall not involve any personal appearance b...

Reviews: