Section 5: Troubleshooting
5- 6
Table 5-4: N -3100 Failure Error Codes (Cont.)
112
1. Turn the monitor off, then on again and enter service
mode.
2. Menu item 5 will be displayed. Save the ICC value as
described in Section 4, Configuration and Service Modes
(paragraph 4.3.5, menu item 5).
159
1. Use the service mode to reset the institutional default
values to factory default values as discussed in Section 4,
Configuration and Service Modes (paragraph 4.3.11, menu
item 20).
2. If the error code still appears, turn off the monitor and
replace the UIF PCB.
177
1. Turn the monitor off, then on again.
2. If stacked and STACKED indicator is not lit, turn all
units in the stack off and then on again.
178, 191
1. Turn the monitor off, then on again.
2. If the error code still appears, use the service mode to
reset the institutional default values to factory default
values as discussed in Section 4, Configuration and
Service Modes (Paragraph 4.3.11, menu item 20).
3. If the error code still appears, turn off the monitor and
replace the UIF PCB.
183
1. Turn the monitor off, then on again.
192, 195, 196
1. Turn the monitor off, then on again.
2. Verify compatibility of software as discussed in Section 4,
Configuration and Service Modes (paragraph 4.3.15, menu
item 29).
3. If the error code still appears, verify compatibility of the
ROMs by calling Nellcor Puritan Bennett’s Technical
Services Department or your local Nellcor Puritan
Bennett representative.