NPB-40 Operation
36
Turn On the NPB-40
Discussion
You must verify that the NPB-40 works properly and is
safe to use. Proper operation of the NPB-40 is verified
each time the NPB-40 is turned on, as described in this
procedure. The verification procedure Power-On
Self-Test (POST) takes 7 to 13 seconds to complete.
Caution: If any indicator or display element does not
display when the NPB-40 is turned on, do not use the
NPB-40. Instead, contact qualified service personnel,
your local Nellcor representative, or Nellcor’s
Technical Services Department.
Note
: Physiological conditions, medical procedures, or
external agents that may interfere with the
NPB-40’s ability to detect and display
measurements, include:
•
dysfunctional hemoglobin
•
arterial dyes
Alarm Silence
Duration
Alarms 30, 60,
90, 120 seconds
30 sec.
Alarm Volume
1 to 10
10
Pulse Beep
Volume
0 to 10
10
Table 2: Parameter Ranges (Continued)
Parameter
Ranges/
Selections
Factory
Defaults
Summary of Contents for Oximax NPB-40
Page 1: ......
Page 8: ...Blank Page...
Page 14: ...Blank Page...
Page 32: ...Blank Page...
Page 38: ...Blank Page...
Page 80: ...Blank Page...
Page 82: ...NPB 40 Operation 74...
Page 98: ...OXIMAX Sensors and Accessories 90 Blank Page...
Page 104: ...Blank Page...
Page 105: ...Performance Considerations NPB 40 97...
Page 106: ...Performance Considerations 98...
Page 116: ...Blank Page...
Page 120: ...Blank Page...
Page 125: ...Menu Structure NPB 40 117 31 November 29 February on a non leap year...
Page 126: ...Blank Page...
Page 132: ...Blank Page...
Page 151: ...Specifications NPB 40 143...
Page 152: ...Specifications 144...
Page 159: ...Index NPB 40 153...
Page 160: ......