Using the Data Port
N-600x Operator’s Manual
113
TxD represents the Transmit Data line, and RxD is the Receive Data
line.
The pin layouts (as viewed from the rear panel of the N-600x) are
illustrated in Figure 6. The conductive shell is connected to earth
ground when connected to a PC or printer.
Figure 6: Data Port Pin Layout
Pins 2, 3, and 5 provide data in RS-232 format.
Pins 1, 4, 9, and 12 provide data in RS-422 format. TxD+ and TxD-
are the differential transmit data pair. RxD+ and RxD- are the
differential receive data pair.
5
Signal Ground (isolated from Earth Ground)
6
AN_SpO
2
(analog saturation output)
7
NC_NO (relay closure nurse call, normally open)
8
NC_NC (relay closure nurse call, normally closed)
9
RxD
-
(RS_422 [
-
] input)
10
Signal Ground (isolated from Earth Ground)
11
Nurse Call (RS
-
232
-
level
-
output)
12
TxD
-
(RS
-
422 [
-
] output)
13
AN_PULSE (analog pulse rate output)
14
AN_PLETH (analog pleth waveform output)
15
NC_COM (relay closure nurse call, common lead)
Table 8: Data Port Pinouts
Pin
Signal Name
Summary of Contents for OXIMAX N-600X Series
Page 1: ......
Page 8: ...Contents vi...
Page 12: ...Safety Information 4 N 600x Operator s Manual...
Page 80: ...Using the Monitor 72 N 600x Operator s Manual...
Page 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Page 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Page 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Page 128: ...Using the Data Port 120 N 600x Operator s Manual...
Page 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Page 146: ...Performance Considerations 138 N 600x Operator s Manual...
Page 168: ...Troubleshooting 160 N 600x Operator s Manual...
Page 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Page 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Page 204: ...Specifications 196 N 600x Operator s Manual...
Page 210: ...Index 202 N 600x Operator s Manual...
Page 211: ......