Troubleshooting
N-550
49
Data Port
Table 10 lists a symptom for data port problems and recommended actions. If
the action requires replacement of the User Interface PCB, refer to
Error Codes
When the N
-
550 detects a defective sensor connected to the N-550, the N-550
displays an error code of “SEN Err.” The sensor should be replaced and the
N-550 power should be cycled.
An error code is displayed when the N-550 detects a non-correctable failure.
Cycling the power clears these errors. Table 11 lists the N-550 error codes and
their meanings. When one of these errors occur, the N-550:
•
sounds a low priority alarm that cannot be silenced except by
power-down
•
stops measurements
Table 10: Data Port Problems
Symptom
Recommended Action
No printout is being
received.
The N-550 is running on battery power. Connect the
N-550 to AC power.
The N-550's baud rate does not match the printer.
Change the baud rate of the N-550 following
instructions in the Operator’s manual.
Check the connection between the data port and the
printer. Refer to
Replace the Front Panel PCB. Refer to
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...