Basic Settings and Calibration
EN
Switching On/Off the compensator:
Press again to change to TILT mode.
Press the button to switch the sensor:
-
On
TILT ON display or
-
Off
TILT
OFF
display.
At the end
of each setting or change, press the
button to
store
the current settings and to
return the display to the normal setting
Angle
Measuring Mode
.
8 Troubleshooting
Note:
If the errors cannot be rectified using one of the listed remedies, the
ET-5
must be examined
by an authorised customer service agent or by NEDO.
Error
Cause
Remedy
The instrument
cannot be switched
on
No power supply
Charge rechargeable
batteries as specified
(see chapter 3.1)
E 01
Counting error, if the reading display
constantly changes during a bearing
Repair required
TOO FAST
The telescope is turned too fast for the
vertical sensor.
Press any button;
except:
and
E 04
Horizontal sensor I Error
Repair required
E 05
Horizontal sensor II Error
Repair required
E 06
Vertical sensor error
Repair required
TILT
The instrument is outside the range of the
TILT sensor. Re-level the theodolite
again. If there is no improvement
TIP:
Switch off the TILT sensor (see 7.1),
the instrument can work without it.
Repair required
ON/OFF
HOLD
0SET
V %