1-12
English.10
Use GammaComp MD Quality Control Software for history management (Constancy test).
Before proper self calibration can be
performed, display should warm-up for a minimum 30 minutes. If the USB sensor is plugged in before the minute warm-up,
a warning will appear on-screen (
Figure S.3
).
NOTE:
Stand-alone calibration can only be performed using the
GretagMacbeth Eye-One Display Sensor.
Refer to the KEY MAP (
Figure A
) when changing settings during the calibration.
1. Plug-in the USB sensor to the USB sensor port (
Figure S.1
).
NOTE:
USB sensor port can only be used for the sensor connection.
2. The calibration menu will open automatically and the sensor will begin to initialize. After initializing, the DARK-
MEASUREMENT message is displayed. If the DARK-MEASUREMENT message is not displayed, please proceed to step 5.
3. Place the USB sensor face down on a non-reflective, opaque surface (ie...desk or a mousepad) to measure the black
level (
Figure S.2
). Press “SELECT”.
4. After the DARK-MEASUREMENT has succeeded (
Figure S.4
), press “SELECT” to continue. If DARK-MEASUREMENT fails
(
Figure S.5
), restart the calibration procedure.
5. Use “+” or “–” to select SELF in the MODE selection (
Figure S.6
). Press “<“ or “>” to move to the next selection. Use “+” or “–”
to select DEFAULT or COPIED WHITE in the WHITE selection (
Figure S.6
). Press “<“ or “>” to move the next selection.
6. Select the target GAMMA: NO CORRECTION, DICOM, LOG LINEAR, PROGRAMMABLE, CUSTOM, and SKIP (
Figure S.6
).
NOTE:
If CUSTOM is selected, the Gamma Value can be manually changed in intervals of 0.1 at 0.5-4.0. If SKIP is
selected, the GAMMA calibration is bypassed reducing calibration time.
7. The procedure will ask for the USB sensor to be placed on the center of the display panel (
Figure S.7
). Tilt the display
panel approximately 5˚ backward and place the USB sensor in the center of the display panel (
Figure S.1
).
NOTE:
Place the USB sensor flat against the LCD to avoid external light contamination.
DO NOT
press the calibrator
against the display panel. Press “SELECT”.
8. Set the target LUMINANCE (cd/m
2
).
Self Calibration
Figure S.1
Sensor Port
Figure S.2
Figure A
KEY MAP
< / >
:
changes from setting to setting
– / +
:
changes the setting selection
(ie...SELF or COPY)
SELECT
:
moves to the next step of calibration
EXIT
:
moves back one step of calibration
USB Sensor
USB icon ( ) is on
the other side