English-21
English
OSD operation without signal input from computer
Without a signal input from the computer, normally a display goes into energy saving mode and cannot show the OSD menu.
MD212G3 and MD211G5 can exit the energy saving mode temporarily and show the OSD menu in the normal operation mode.
Therefore it is possible to change settings in the OSD or perform a DICOM conformance test without changing the computer’s
power management settings.
Follow the procedures below to show the OSD menus.
1. Confi rm that the display is in the energy saving mode and the LED next to the power switch is lit orange. If the LED is off,
please push the power switch to turn on the monitor.
2. Hold down the (+) plus button and the (-) minus button simultaneously for over 2 seconds.
The “INFORMATION” OSD main menu opens on the screen. The LED color changes from orange to green.
3. Switch EDID, input source, confi guration or DICOM conformance test, referring to this manual.
Changing the EDID settings (see page 16)
Changing the Display Mode* (see page 16)
Input source switching (see page 17)
Changing the confi guration (see page 18)
Performing the DICOM conformance test (see page 19)
4. If the display enters the energy saving mode during this operation, please start from the beginning again.
NOTE:
The display being turned on to the normal operation mode by the operation as described above is only temporary.
This mode is left after any of the following conditions are met:
1. 3 hours have passed.
2. The display power has been turned off.
3. A DICOM conformance test has been performed.
*: MD211G5 only.