DC-3000C SERIES ULTRASONIC THICKNESS GAUGE
- 12 -
3. Calculation error caused by the sound velocity set in the instrument is
different from that of the actual material.
In order to eliminate the possible error, please use following calibration
method:
Measure the test piece with known thickness. Get a measurement value,
Press
into menu-
“SETTING”-“PROBE CALIBRATION”,
Pressing
or
to adjust the measured value as the same as the
value of sample that is measured,
Press
to confirm.
Notice: Please notice that calibration range is within ±0.25mm. If it is out
of range, which means users should replace a new probe.
【
8.6
】
Memory (DC-3020C)
Press
into the menu, and then go into “Memory”, the screen will
display
1 Memory unit
2 Memory read
3 Delete memory
4 Data transfer
Press
or
into the selected item, press
to confirm.
8.6.1
Memory unit
The gauge has a memory capacity of 5,200 measurements. The memory
location was composed by alphabet A-Z +
location number. There are