2030434-006 Rev. E
EasyOne-CS Field Service Manual
Page 22
13.2
Environment
1.
Windows PC with installed and working EasyOne-CS connection (Serial cradle, USB cradle or
Screen-Connector)
2.
Installed EasyWare greater or equal V02.04
3.
Torx T8 / Allen key /hex key screw driver depending on serial number
4.
Electrostatic save working environment for the device disassembly
13.3
Preparation (save database)
To make sure that no data is lost/altered, the internal EasyOne-CS database should be saved first.
Please follow the below steps.
1.
Start EasyWare by the link on the desktop
2.
Switch EasyOne-CS on and press key sequence 1-3-5 when the below shown error screen
appears. Enter the date & time if requested. Then the normal main menu should be shown.
ERROR
Device Selftest
Error #25
Please contact
service
3.
Connect the EasyOne-CS to the PC (Device Connected is shown in the status line of EasyWare/
“PC interface active” on the EasyOne-CS screen)
4.
Go to menu: File/Backup Device DB… And the following progress bar appears
5.
The EasyOne-CS backup file will be written (takes several minutes) and the success is indicated
by the following dialog
6.
The backup file of the form EZ1BackupXXXXX_DateTime.txt
- XXXXX: device serial number
- Date: YearMonthDay
-Time: HourMinute
(e.g.
EZ1Backup40126_200506071607.txt)
is located in the folder where EasyWare has been
started e.g. C:\programmfiles\ndd Medizintechnik\EasyWare
Summary of Contents for EasyOne-CS
Page 2: ......