2030434-006 Rev. E
EasyOne-CS Field Service Manual
Page 16
11
EasyOne-CS Device Software Update
11.1
Introduction
The internal software of EasyOne-CS is stored in a Flash EPROM. It is therefore possible to update the
internal software. The following procedure describes how this update is executed.
11.2
Validated Upgrade Steps
The following upgrade steps are validated:
Old Version
New Version
Item Number
EasyOne-CS device software
CRC
V04.01.00.00
V04.03.00.00
2033231-008
GE-EOV04030000-P-EU_Resources.hex
with EU language package
0xF3D7
11.3
Preparation
The following items are required:
2033231-008.zip (upgrade package with EasyFlash and EasyOne-CS device software V4.3)
ndd Screen-Connector cable 2030431-003.
PC program EasyFlash, V2.4.12 or higher.
EasyOne-CS device software for the download (*.hex).
PC with Windows 2000, XP or Vista (all 32 Bit).
The following installation description presumes that you have basic knowledge of your PC operating
system.
11.4
Unpacking of Upgrade Package
After getting upgrade package 2033231-008.zip the zip archive can be extracted to a folder of your
choice, e.g. to C:\nddmed\2033231-008.
11.5
Installation of EasyFlash PC Software
The PC software EasyFlash is needed to update EasyOne-CS device software:
1.
Start “C:\nddmed\2033231-008\EasyFlash\Setup.exe” to install the EasyFlash software.