Turn on control unit using the side toggle switch. One 
of the green or red LEDs will come on. 
 

Set Target Range 

 

The therapist needs to set the correct upper and 
lower limits.  Make sure the two dials are in the 
following default positions: 

 

Upper limit dial fully clockwise (maximum 
position)  

 

Lower limit dial fully anti-clockwise (minimum 
position). 

 
Firstly, program the lower limit. Turn this Lower Limit 
dial clockwise until you hear a sound from the 
speaker simultaneously the target weight is being put 
through the insole.

 

 

Now get the upper limit by turning this dial 
anticlockwise, until you hear a sound from the 
speaker while weight is being put through the insole. 
Finesse this dial a little (by turning clockwise) to 
create your window.

 

 

In summary, when the 

client’s

 partial weight on the 

insole is between the lower and upper limit values, 
the client will hear no sound. If the weight shift is 
outside these two limits, then the client will receive 
feedback. The sound changes and increases in pitch 
the further away the client gets from the target range. 

 

Before using the headphones turn the volume control 
back to minimum then slowly increase until the sound 
is comfortable for the user.  

 

FAQs 
 

Q. Can I just use the heel sensor on its own? 
 
A.  Yes, If you are just using one sensor you will not 
need the adaptor. 
 
Q. Can the unit be set so user gets feedback only 
when they reach the pre-set target? 
 
A. Yes, turn lower limit dial fully anti clockwise and 
adjust upper limit dial until a sound is generated 
when target weight is being put through the insole. 

 

  
 
 
 

 
 
 
 

 
 
 

 
 
 
 

 
 
 
 

 
 
 
 

 
 
 
 

 
 
 
 

 
 
 
 

 
 
 
 

 
 
 
 

 
 
 
 
Problems? 

 

Please do not hesitate to contact nCounters via 

email at: 

[email protected] 

or via our enquiry page 

ncountersonline.com

 