7 Reference
7.2 mV (oxidation-reduction potential [ORP]) measurement
D-52/53/54/55
183
●
ORP standard solution
There are two kinds of standards substances. Under
normal circumstances, it is sufficient to use only that
type of substance which is closest to the measured
value.
Indicated value of ORP standard solution at
various temperatures (mV)
temp.
(ºC)
160 - 22
Phthalic-acid
ch
quinhydrone
160 - 51
Neutral
pho
quinhydrone
5
+274.2
+111.9
10
+270.9
+106.9
15
+266.8
+101.0
20
+262.5
+95.0
25
+257.6
+89.0
30
+253.5
+82.7
35
+248.6
+76.2
40
+243.6
+69.0
Summary of Contents for D-25
Page 2: ......
Page 11: ...CONTENTS VIII HORIBA ...
Page 31: ...1 Overview of the Meter 1 9 Using the protective cap D 53 54 55 20 HORIBA ...
Page 75: ...2 Taking Measurements 2 9 Dissolved oxygen DO measurement D 55 64 HORIBA ...
Page 139: ...4 RS 232C communications 4 5 Communication example using the HyperTerminal 128 HORIBA ...