General Safety Guidelines

1. Use the Stand-Up Lift only for its intended purpose, in accordance with the legislation for medi-

cal devices, the regulations for labor protection and accident prevention, as well as the generally 

recognized rules of technology.
2. Note that the Stand-Up Lift is a medical device, therefore the user is required to comply with the 

directive on the use of medical devices.
3. The requirements for the electrical installation of the room or area where the Stand-Up Lift is 

used must meet the current state of the technology.
4. Using the Stand-Up Lift only after being trained for its handling.
5. Before using the device, please read the complete User Manual in order to avoid damage due 

to improper handling or exposure to risks. The User Manual contains important information and 

notes required for the Stand-Up Lift use.
6. Use the Stand-Up Lift in accordance with this User Manual only. Keep the User Manual for future 

reference in case of questions. Join this User Manual to the Stand-Up Lift in case of change of 

ownership.
7. Before any use, it must be ascertained whether the Stand-Up Lift and its accessories are in wor-

king order and in impeccable conditions.
8. Before using the Stand-Up Lift with other medical and non-medical devices, check that the 

combination of these products is allowed and can be used together safely.
9. Assembly, commissioning, maintenance and repair of the Stand-Up Lift should be entrusted 

only to specialists.
10. It is up to the user or operator to ensure (through adequate measures and instructions) that 

mechanical stress of the charging cord (through bending, tension, shear, crushing) is excluded 

during loading or cleaning of the part. This also applies to the electrical cables of other devices 

used with the Stand-Up Lift.
11. Respect the activation duration and the maximum weight capacity. These values must not be 

exceeded, otherwise the Stand-Up Lift safe operation is no longer guaranteed.
12. Do not expose the Stand-Up Lift to direct sunlight or heat.
13. Ensure that no moisture enters the electrical system.
14. Avoid mechanical stress on electrical cords. Pull, bend or crush the cords may damage them.
15. Charge batteries in a well ventilated area.
16. Electromagnetic interferences can not be excluded from the Stand-Up Lift and other devices. 

If there is a risk of such interference, the source of interference must be removed or the Stand-Up 

Lift should not be used.
17. Disruptions caused by the use of portable communication devices can not be completely 

excluded. This is why a safe distance of at least 3 meters must be kept in order to ensure the 

functioning of the Stand-Up Lift safely.
18. Do not leave unattended children stand near the Stand-Up Lift.
19. The Stand-Up Lift should not be used as soon as abnormal noises, damages or any other mal-

functions occur. In this case, do not connect the wall charger to the Stand-Up Lift and inform us.
20. If damaged or defective, the Stand-Up Lift should not be used and should not be connected to 

the direct current. Inform the dealer to remedy the default or failure.

14

User Manual - Stand-Up Lift: EASYLEV 5 Range

Manufactured by NAUSICAA Médical S.A.S. / Approved by Ghizlane Labrosse (Biomedical Engineer)

Summary of Contents for EASYLEV 5 Series

Page 1: ...MEDICAL DEVICE CONFORM TO STANDARD NF EN ISO 10535 2007 NAUSICAA Medical is certified ISO 13485 www nausicaa medical eu User Manual Stand Up Lift EASYLEV 5 Range Manufactured by NAUSICAA M dical S A...

Page 2: ...eling General Safety Guidelines Technicals Dimensionals Characteristics EASYLEV5 EASYLEV5 with mechanical opening legs EASYLEV5 PLUS with electrical opening legs Technicals Dimensionals Characteristic...

Page 3: ...ir a toilet or a wheelchair to be transferred while being active 1 The stand up lift has already been inspected at the factory to ensure that it is free from defects and that nothing is missing Nevert...

Page 4: ...deformation on the hooks WARNING To preserve the inserts do not overtighten Assembly is done with the brakes locked Assembling Instructions EASYLEV5 Step 1 Take the device out of the box The device i...

Page 5: ...no wear or deformation on the hooks Assembling Instructions EASYLEV5 with mechanical opening legs EASYLEV5 PLUS with electrical opening legs Step 1 Take the device out of the box The device is ready f...

Page 6: ...the parking lever pointing upwards Castor Released When locking the castors always insure that both castors are locked If only one castor is locked on a sloping floor the unlocked wheel will roll down...

Page 7: ...tonicity patients How to properly use a stand up lift 1 Low tonicity patients shall be transferred in a semi seated position Instructions for Use Your stand is designed to lift people so do not use it...

Page 8: ...stand up lift 2 Medium tonicity patients shall be transferred in a standing position Likewise the patient s shins must be in contact with the shin block at all times In all cases the caregiver must b...

Page 9: ...tatus of the control box and the battery Check that the actuator is properly secured Check that the battery is charged Stand up Lifts Use 7 Control Box 2 Functions Control Box 4 Functions User Manual...

Page 10: ...rned are EASYLEV5 EASYLEV5 with me chanical opening legs The products concerned are EASYLEV5 EASYLEV5 with mechanical opening legs The products concerned are EASYLEV5 Plus with electrical opening legs...

Page 11: ...oo much Standing up for PATIENT WITH CHEST TONICITY INSTRUCTIONS FOR USE Position the bottom of the backrest in the lumbar region in order that the ears of the attachment system are on either side of...

Page 12: ...ty An assessment of the patient capacity is required before using a stand up lift and more precisely to make sure that the patient can bear weight on at least one leg and has sufficient tonicity in th...

Page 13: ...66 51 50 47 Web www nausicaa medical com 12 All e du Piot ZA P le Actif 30660 Gallargues le Montueux FRANCE Tel 33 4 66 51 50 80 Fax 33 4 66 51 50 47 Web www nausicaa medical com 12 All e du Piot ZA...

Page 14: ...ANCE A POORL Y MAINTAINED BATTERY DETERIORATES SAFETY INSTRUCTIONS BEFORE EACH USE Verify the presence and conditions of safety devices all axis and their constrained fixations pins clips and bolts an...

Page 15: ...Labelling 13 For all models of the EASYLEV 5 Range User Manual Stand Up Lift EASYLEV 5 Range Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer 3 4 2 1 5 7 8...

Page 16: ...ensure through adequate measures and instructions that mechanical stress of the charging cord through bending tension shear crushing is excluded during loading or cleaning of the part This also applie...

Page 17: ...675 Maximum height Low handles Position 1 1475 Position 2 1500 Position 3 1525 Handles high Position 1 1500 Position 2 1520 Position 3 1545 Low handles Position 1 1490 Position 2 1515 Position 3 1540...

Page 18: ...eight 455 420 4 distance from the device to the maximum range 565 530 5 rounding diameter 900 1105 1 total length 840 995 9 total width closed feet 1210 1260 1 Maximum charge 180kg 180kg 18 Total weig...

Page 19: ...9 cm Electronic protection when overloaded Protections types T l commande IP56 Remote control IP54 Control box IP65 Battery pack IP65 Actuator IP55 Stop limit switch Box in ABS Handcontrol with spiral...

Page 20: ...s le v rin va s arr ter et une enclume s affichera Il sera alors poursuivre le fonctionnement de l appareil ompl te e toutes les charges incompl tes de la batterie inf rieures fre correspond au nombr...

Page 21: ...steam use with a microfiber mop Do not direct steam directly on the electrical boxes MONTHLY MAINTENANCE Check that the device does not present any visible damages Check that no part is missing Check...

Page 22: ...rt check Sling fixation hooks check State of the frame and of the painting check Additional checks for models with opening base legs Foot pedal or opening legs actuator check Spreaded legs system chec...

Page 23: ...vot of wheels Sub patellar support slide ECP Pedal or Cylinder Assembly Foot Spacer System Assemblies of the right and left feet on Embase 8 Periodic lubrication Use a 3 in 1 product or similar Period...

Page 24: ...t runner ECP Pedal or Cylinder Assembly Foot Spacer System Assemblies of the right and left feet on Embase 8 22 Periodic lubrication Use a 3 in 1 product or similar EASYLEV5 PLUS with electrical openi...

Page 25: ...ION SLING FOR EASYLEV 5 ESL5 STR PF 13 FRONT AND SIDE WHEELS FOR EASYLEV AND WAYUP 5 WITH FIXED BASE WP5 RJND50 PF 14 REAR WHEELS FOR EASYLEV AND WAYUP 5 WITH FIXED BASE WP5 RJNFRD100 PF 15 3000N LIFT...

Page 26: ...F 7 RIGHT FOOT FOR EASYLEV AND WAYUP 5 WP5 EP PD PF 8 LEFT FOOT FOR EASYLEV AND WAYUP 5 WP5 EP PG PF 9 REMOVABLE PLATFORM FOR EASYLEV AND WAYUP 5 WP5 EP PLA PF 10 HEEL BLOCK FOR RANGE 5 WP5 CLT PF 11...

Page 27: ...ECP EMB PF 8 RIGHT FOOT FOR EASYLEV AND WAYUP 5 WP5 EP PD PF 9 LEFT FOOT FOR EASYLEV AND WAYUP 5 WP5 EP PG PF 10 REMOVABLE PLATFORM FOR EASYLEV AND WAYUP 5 WP5 EP PLA PF 11 HEEL BLOCK FOR RANGE 5 WP5...

Page 28: ...OULEUR DE CORDON SECTEUR UMEC VB 26 User Manual Stand Up Lift EASYLEV 5 Range Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer 2 2 7 2 6 2 8 2 4 2 5 2 3 2 9 1 3...

Page 29: ...LEUR DE CORDON SECTEUR UMEC VB 27 User Manual Stand Up Lift EASYLEV 5 Range Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer 2 1 2 7 2 6 2 8 2 4 2 5 2 3 2 9 2 1...

Page 30: ...X III 28 User Manual Stand Up Lift EASYLEV 5 Range Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer 2 1 5 3 4 NUMBER DESIGNATION REFERENCE 1 2 FUNCTIONS REMOTE...

Page 31: ...Stand Up Lift EASYLEV 5 Range Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer 5 4 1 2 3 6 NUMBER DESIGNATION REFERENCE 0 COMPLETE CONTINUED EMERGENCY LOWERING...

Page 32: ...CHAPE D EMBASE DE VERIN D ECARTEMENT DES PIEDS 3000N GRIS VS VRN ECP3 CEMB 3 CORDON SPIRALE POUR VERIN D ECARTEMENT DES PIEDS 3000N GRIS VSVRN ECP3 CRD Opening Legs Jack 3 000 Newtons NUMBER DESIGNATI...

Page 33: ...The batteries are discharged or worn out Charge the device or change the batteries The weight lifted by the device is too important a weight symbol is displayed on the screen Decrease the weight lift...

Page 34: ...32 User Manual Stand Up Lift EASYLEV 5 Range Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer...

Page 35: ...does not apply if the claims are consecutive to accident misuse or neglect of the unit by the end customer shipping performed without adequate protection alteration or transformation not validated by...

Page 36: ...e 33 0 466 717 184 Fax 33 0 466 532 305 Email export nausicaa medical com Website www nausicaa medical eu Manufactured by NAUSICAA M dical S A S 12 All e du Piot ZA P le Actif 30660 Gallargues le Mont...