5.7
Analyze Slow Phase Velocity (SPV) Measurements
A. Click the
Review
tab and select a Caloric or other test to analyze nystagmus.
B. Press
F12 Begin
to measure the slow phases and plot them on the lower chart. The number of beats analyzed displays
in the Information area.
C. Use the function keys to work with the results:
Function Keys, during SPV Analysis
F2
Reanalyze
Clear previous analysis and begin a new analysis.
F4
Pods - Butterfly
Display pod (summary) and butterfly view of caloric ana-
lysis.
F5
Locate Peak
Identifies the second in time with the fastest average
value.
F6
Set Peak
Computes the average of the highest SPVs in the 10-
second window at the current cursor location.
F7
Delete
Delete the SPV beat measurement.
F8
Insert
Accept a new SPV value for the beat.
F9
Close
Close Analysis
F10
Previous Beat
Move to previous beat.
F11
Next Beat
Move to next beat.
F12
Save
Save analysis and return to Review mode.
D. (Optional) Measure the slope of a beat manually.
SPV Analysis
1. Select a beat (A):
–
Click and drag the scroll bar or
press the
áâ
arrow keys to
move through the tracing (B).
–
Click
F10 Previous Beat
and
F11 Next Beat
to move
through the tracing.
–
Click on a beat in the lower
chart or click in the upper chart
to move to that section on the
tracing.
2. Click
F7 Delete
to delete artifactual
beats
36
Otometrics - ICS Chartr 200 VNG/ENG
5 Review and Analyze Collected Data
Summary of Contents for Otometrics ICS Chartr 200 VNG/ENG
Page 4: ......