4.3
Check the Range
A range sensor provides a continuous display of patient-to-light bar distance. The acceptable range during testing is 3 feet 8
inches to 4 feet 4 inches (111.8 cm to 132.1 cm). If a patient is “out of range” for more than 10% of a test, a message dis-
plays at the end of the test.
A. Click the
New Test
tab on the Main Window and select a protocol.
B. Position the patient in front of the light bar.
C. Click
F6 Range
to activate the range sensor.
Range Dialog Box
Range:
•
Distance – in feet / inches and centimeters
•
Over Range
•
Under Range
During a test, range status updates every 3 seconds and dis-
plays in the Information Area of the Main Window.
D. Click
OK
to close the dialog and exit the Interpretation Assistant.
E. Go to
.
Otometrics - ICS Chartr 200 VNG/ENG
21