Contraindications, Warnings, & Cautions
Contraindications
The Brain Monitor & Embla Dx Series Amplifiers shall NOT be used in the following conditions:
Do
NOT
operate the system in the presence of flammable anesthetics.
Check areas of use to avoid using the system in the presence of flammable gases.
To ensure the validity of signals, do not operate the device near any sources of electromagnetic
interference.
Natus
systems are not AP or APG rated.
DO NOT USE
a Natus system in the presence of a
flammable anesthetic mixture with air, oxygen, or nitrous oxide.
The
Brain Monitor & Embla Dx Series Amplifiers
are
NOT
designed to work with
defibrillators. The Brain Monitor & Embla Dx systems could be damaged when used with this
device.
When using the Brain Monitor & Embla Dx Series Amplifiers in the vicinity of electrosurgical
units (cautery devices) it is recommended to connect to the acquisition PC using the TCP/IP
interface.
Operation of this equipment with input signals in excess of a range of +/-10mV may cause
incorrect results.
Warnings and Cautions
General Warnings
NOTE:
It is recommended that all data be stored using redundant storage capabilities. This can
help to minimize data loss in the event of a failure of the primary drive.
This equipment/system is intended for use by Healthcare professionals
ONLY
. Please read this
section before installing any of the hardware. Refer to this section when you operate, transport,
store, or re-install the system.
Only use the Brain Monitor & Embla Dx Series Amplifiers in conjunction with approved devices
and accessories.
Use of cables other than those specified or sold by the manufacturer on the equipment, may
result in increased emissions or decreased immunity of the equipment and may cause the
system to be non-compliant with the requirements of IEC 60601-1-2:2007
Never use equipment that has parts missing or equipment that might contain loose parts inside
of it (that is, inside an enclosed portion of the equipment). If you suspect a piece of equipment
has missing or loose parts, contact Natus. Routinely inspect system cables and components for
regular wear and tear.