Natus ErgoJust LTM and ErgoJust ICU Cart Instructions For Use
4-12
July 6, 2020
Fixed-zoom video option power cables connections -
Natus ErgoJust ICU
Ensure that the ISO or UPS is disconnected (unplugged) from the MAINS power source before connecting power cables to
the system components.
See the Fixed-zoom video option cable routing diagram on the next page.
Figure 1: Fixed-zoom video option power cable connections.
Legend
Description
Legend
Description
A
Do not connect power to these two inputs
G
MAINS
B
12V-
H
Power Supply
C
12V+
I
ISO (shown) or UPS
D
To RD, TD, GND camera control cables
J
Natus Amplifier Base Unit
E
Fixed-zoom Video Camera
K
All-in-One Computer
F
IR Light
L
Connectors and device styles may vary from those shown.