With Aurical® HIT you receive a set of color-coded battery simulators, which are used to power the hearing instru-
ment. They are also used as probes for measuring the power consumption.
Color code
Size
IEC
ANSI
Red
5
PR63
7012ZD
Yellow
10
PR70
7005ZD
Brown
312
PR41
7002ZD
Orange
13
PR48
7000ZD
Blue
675
PR44
7003ZD
2. Insert the mini-jack connector of the battery simulator in the bat-
tery simulator socket in the test chamber.
Aurical® HIT automatically detects the battery simulator.
Warning
•
When you have connected the battery simulator,
make sure that it does not touch other metal parts as this may
short-circuit the system.
5.4
Coupler adapters
Adapters for use with the coupler are snapped onto the coupler cavity.
•
HA-2 (BTE)
Traditional BTE hearing instruments
•
HA-1 (ITE, RIE, thin-tube)
and
To fix the hearing instrument on the adapter, snap the adapter off the coupler cavity, and attach the hearing instrument to
the adapter outside the test chamber.
Aurical HIT
11
5 Testing hearing instruments
Summary of Contents for Aurical HIT